Day-to-day reliability of basal heart rate and short-term and ultra short-term heart rate variability assessment by the Equivital eq02+ LifeMonitor in US Army soldiers.

作者信息

Chapman Christopher L, Schafer E A, Potter A W, Lavoie E M, Roberts B M, Castellani J W, Friedl K E, Looney D P

机构信息

US Army Research Institute of Environmental Medicine, Natick, Massachusetts, USA.

Oak Ridge Institute for Science and Education, Oak Ridge, Tennessee, USA.

出版信息

BMJ Mil Health. 2024 Jul 14. doi: 10.1136/military-2024-002687.

Abstract

INTRODUCTION

The present study determined the (1) day-to-day reliability of basal heart rate (HR) and HR variability (HRV) measured by the Equivital eq02+ LifeMonitor and (2) agreement of ultra short-term HRV compared with short-term HRV.

METHODS

Twenty-three active-duty US Army Soldiers (5 females, 18 males) completed two experimental visits separated by >48 hours with restrictions consistent with basal monitoring (eg, exercise, dietary), with measurements after supine rest at minutes 20-21 (ultra short-term) and minutes 20-25 (short-term). HRV was assessed as the SD of R-R intervals (SDNN) and the square root of the mean squared differences between consecutive R-R intervals (RMSSD).

RESULTS

The day-to-day reliability (intraclass correlation coefficient (ICC)) using linear-mixed model approach was good for HR (0.849, 95% CI: 0.689 to 0.933) and RMSSD (ICC: 0.823, 95% CI: 0.623 to 0.920). SDNN had moderate day-to-day reliability with greater variation (ICC: 0.689, 95% CI: 0.428 to 0.858). The reliability of RMSSD was slightly improved when considering the effect of respiration (ICC: 0.821, 95% CI: 0.672 to 0.944). There was no bias for HR measured for 1 min versus 5 min (p=0.511). For 1 min measurements versus 5 min, there was a very modest mean bias of -4 ms for SDNN and -1 ms for RMSSD (p≤0.023).

CONCLUSION

When preceded by a 20 min stabilisation period using restrictions consistent with basal monitoring and measuring respiration, military personnel can rely on the eq02+ for basal HR and RMSSD monitoring but should be more cautious using SDNN. These data also support using ultra short-term measurements when following these procedures.

摘要

引言

本研究确定了(1)使用Equivital eq02+生命监测仪测量的基础心率(HR)和心率变异性(HRV)的每日可靠性,以及(2)超短期HRV与短期HRV的一致性。

方法

23名美国现役陆军士兵(5名女性,18名男性)完成了两次实验性访视,间隔超过48小时,且有与基础监测一致的限制条件(如运动、饮食),在仰卧休息20-21分钟(超短期)和20-25分钟(短期)后进行测量。HRV通过R-R间期标准差(SDNN)和连续R-R间期平方差的平方根(RMSSD)进行评估。

结果

采用线性混合模型方法得出的每日可靠性(组内相关系数(ICC)),对于HR(0.849,95%CI:0.689至0.933)和RMSSD(ICC:0.823,95%CI:0.623至0.920)良好。SDNN的每日可靠性中等,变异较大(ICC:0.689,95%CI:0.428至0.858)。考虑呼吸影响时,RMSSD的可靠性略有提高(ICC:0.821,95%CI:0.672至0.944)。1分钟与5分钟测量的HR无偏差(p=0.511)。对于1分钟与5分钟测量,SDNN的平均偏差非常小,为-4毫秒,RMSSD为-1毫秒(p≤0.023)。

结论

当在采用与基础监测一致的限制条件并测量呼吸的20分钟稳定期之后,军事人员可以依靠eq02+进行基础HR和RMSSD监测,但使用SDNN时应更加谨慎。这些数据也支持在遵循这些程序时使用超短期测量。