J Refract Surg. 2024 Jul;40(7):e499-e505. doi: 10.3928/1081597X-20240522-01. Epub 2024 Jul 1.
To assess the effects of intraocular lens (IOL) decentration and tilt, as well as age, on postoperative visual function (corrected distance visual acuity [CDVA] and contrast sensitivity) by comparing an extended depth-of-focus IOL using higher order aspheric optics against a monofocal IOL from the same platform.
This retrospective observational study targeted patients without other eye diseases who underwent surgery to implant the Tecnis Eyhance OptiBlue or the monofocal IOL Tecnis OptiBlue 1-Piece (J&J Vision) during cataract surgery from November 2021 to December 2022. The effects of age, axial length, IOL decentration, tilt, and corneal higher order aberrations (HOAs) on the postoperative 5 m CDVA and area under log contrast sensitivity function (AULCSF) under photopic and scotopic conditions were evaluated within 3 months of surgery.
No significant difference was found in postoperative CDVA between the Tecnis Eyhance OptiBlue group (n = 61 eyes) and the Tecnis OptiBlue 1-Piece group (n = 35 eyes), but AULCSF was significantly better in the Tecnis Eyhance OptiBlue group for photopic (1.58 ± 0.13 vs 1.46 ± 0.18; = .002) and scotopic (1.71 ± 0.11 vs 1.59 ± 0.19; = .002) eyes. Multivariate analysis showed a negative correlation between AULCSF and IOL decentration and age in the Tecnis Eyhance OptiBlue group ( < .01), with no significant correlation with tilt, axial length, and corneal HOAs.
The Tecnis Eyhance OptiBlue yielded significantly better contrast sensitivity under photopic and scotopic conditions than the Tecnis OptiBlue 1-Piece. However, it is important to consider the effects of IOL decentration and age when evaluating the contrast sensitivity of the Tecnis Eyhance OptiBlue. .
通过比较具有更高阶像差非球面光学设计的扩展景深人工晶状体(IOL)与同平台的单焦点 IOL,评估 IOL 偏心和倾斜以及年龄对术后视觉功能(矫正远距视力[CDVA]和对比敏感度)的影响。
这是一项回顾性观察研究,纳入了 2021 年 11 月至 2022 年 12 月期间接受白内障手术并植入 Tecnis Eyhance OptiBlue 或单焦点 IOL Tecnis OptiBlue 1-Piece(J&J Vision)的无其他眼部疾病的患者。评估术后 3 个月内 IOL 偏心、倾斜、年龄、眼轴长度、角膜高阶像差(HOA)对术后 5 m CDVA 和明、暗光条件下对数对比敏感度功能下面积(AULCSF)的影响。
Tecnis Eyhance OptiBlue 组(n = 61 眼)和 Tecnis OptiBlue 1-Piece 组(n = 35 眼)术后 CDVA 无显著差异,但 Tecnis Eyhance OptiBlue 组明、暗光眼的 AULCSF 均显著更好(分别为 1.58 ± 0.13 与 1.46 ± 0.18; =.002 和 1.71 ± 0.11 与 1.59 ± 0.19; =.002)。多变量分析显示,Tecnis Eyhance OptiBlue 组 AULCSF 与 IOL 偏心和年龄呈负相关( <.01),与倾斜、眼轴长度和角膜 HOA 无显著相关性。
与 Tecnis OptiBlue 1-Piece 相比,Tecnis Eyhance OptiBlue 在明、暗光条件下的对比敏感度显著提高。然而,在评估 Tecnis Eyhance OptiBlue 的对比敏感度时,需要考虑 IOL 偏心和年龄的影响。