Zhou Qi
Cardiovascular Medicine Department, Hebei Petro China Central Hospital, Langfang, Hebei Province, China.
Blood Press Monit. 2025 Feb 1;30(1):36-39. doi: 10.1097/MBP.0000000000000717. Epub 2024 Jul 16.
The aim of this study was to evaluate the accuracy of the DBP-8278B Wrist-type Fully Automatic Blood Pressure Monitor in the general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018/AMD 1 : 2020). The test monitor was used to measure SBP and DBP and pulse rate in the wrists using the oscillometric method. According to this protocol using the same-arm sequential blood pressure measurement method, subjects were recruited to fulfill the age, sex, blood pressure, and cuff distribution criteria of the Universal Standard. A single cuff for wrist sizes of 13.5-21.5 cm was used. A total of 85 subjects were included in this study. According to Criterion 1, the mean difference in SBP between the test device and the reference device was 0.89 mmHg, with a SD of 7.76 mmHg. The mean difference of DBP was -1.44 mmHg, with a SD of 6.23 mmHg. The mean difference between SBP and DBP was less than 5.0 mmHg, and the SD was less than 8.0 mmHg, which met the requirements. According to Criterion 2, the SD of the mean differences of SBP between the test device and the reference device per subject was 5.65 mmHg, which was less than 6.88 mmHg and met the requirements. The SD of the mean differences of DBP between the test device and the reference device per subject was 5.61 mmHg, which was less than 6.78 mmHg and met the requirements. The DBP-8278B Wrist-type Fully Automatic Blood Pressure Monitor met the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018/AMD 1 : 2020). It can be recommended for home and clinical use.
本研究旨在根据美国医疗器械促进协会(AAMI)/欧洲高血压学会(ESH)/国际标准化组织(ISO)通用标准(ISO 81060-2:2018/AMD 1:2020),评估DBP-8278B腕式全自动血压计在普通人群中的准确性。使用该测试血压计通过示波法测量手腕处的收缩压(SBP)、舒张压(DBP)和脉搏率。根据此采用同侧连续血压测量方法的方案,招募符合通用标准的年龄、性别、血压和袖带尺寸分布标准的受试者。使用一个适合手腕周长为13.5 - 21.5厘米的袖带。本研究共纳入85名受试者。根据标准1,测试设备与参考设备之间SBP的平均差值为0.89毫米汞柱,标准差为7.76毫米汞柱。DBP的平均差值为 -1.44毫米汞柱,标准差为6.23毫米汞柱。SBP和DBP的平均差值均小于5.0毫米汞柱,标准差均小于8.0毫米汞柱,符合要求。根据标准2,每个受试者测试设备与参考设备之间SBP平均差值的标准差为5.65毫米汞柱,小于6.88毫米汞柱,符合要求。每个受试者测试设备与参考设备之间DBP平均差值的标准差为5.61毫米汞柱,小于6.78毫米汞柱,符合要求。DBP-8278B腕式全自动血压计符合AAMI/ESH/ISO通用标准(ISO 81060-2:2018/AMD 1:2020)的要求。可推荐用于家庭和临床使用。