Garrett J S, Lanzer P, Jaschke W, Botvinick E, Sievers R, Higgins C B, Lipton M J
Am J Cardiol. 1985 Oct 1;56(10):657-61. doi: 10.1016/0002-9149(85)91030-6.
High-speed computed transmission tomography (cine CT) is a new noninvasive technique that may be useful for the rapid, accurate quantitation of cardiac function. The capability of cine CT to assess cardiac output was examined 10 mongrel dogs as an initial step in validating this method. After the dogs were anesthetized, femoral arterial pressure and pulmonary artery thermodilution catheters were inserted. After intravenous injection of a bolus of contrast medium, cine CT scans were performed in the flow mode, in which 50-ms scans were triggered electrocardiographically at end-diastole at 8 levels during sequential cardiac cycles. Scans and thermodilution measurements of cardiac output were obtained at rest and during altered hemodynamic states induced by separate infusions of dobutamine and phenylephrine. Time-density analysis was performed over the left ventricular cavity and curves were fitted to the CT flow data by gamma-variate analysis. Using the Stewart-Hamilton equation established for indicator dilution techniques, the cardiac output was calculated. The results established a direct linear correlation (r = 0.92) between cine CT cardiac output and thermodilution cardiac output over a wide range of cardiac outputs (1.5 to 6.3 liters/min). This study demonstrates that cine CT can provide a reliable estimate of cardiac output noninvasively using contrast medium as an indicator in dogs.
高速计算机断层扫描传输技术(电影CT)是一种新的无创技术,可能有助于快速、准确地定量评估心功能。作为验证该方法的第一步,对10只杂种犬进行了电影CT评估心输出量能力的研究。犬麻醉后,插入股动脉压力导管和肺动脉热稀释导管。静脉注射一剂造影剂后,以血流模式进行电影CT扫描,在连续心动周期中,于舒张末期在8个层面通过心电图触发进行50毫秒的扫描。在静息状态以及分别输注多巴酚丁胺和去氧肾上腺素引起血流动力学状态改变时,获取扫描图像和心输出量的热稀释测量值。对左心室腔进行时间-密度分析,并通过伽马变量分析将曲线拟合到CT血流数据。使用为指示剂稀释技术建立的斯图尔特-汉密尔顿方程计算心输出量。结果表明,在较宽的心输出量范围(1.5至6.3升/分钟)内,电影CT测得的心输出量与热稀释法测得的心输出量之间存在直接线性相关性(r = 0.92)。本研究表明,电影CT能够以造影剂作为指示剂,无创地提供犬心输出量的可靠估计值。