Monitoring of changes in skeletal muscle oxygenation during exercise has increased in recent years. Tissue oxygenation, which is related to fatigue and muscle hypertrophy, is often measured using near-infrared spectroscopy (NIRS).
This study aimed to determine the test-retest reliability of a non-portable NIRS (NIRO200Nx) during the full-squat exercise and recovery in young healthy men.
Twenty-five male participants (21.8 ± 2.6 years) were recruited for this original research. Each participant completed an 8-repetition test with a load that elicited a velocity of 1 m·s. The test was conducted twice, with a 48-hour washout period between sessions.
The NIRS measured the changes of oxygenated-Hemoglobin (OHb), deoxygenated-Hemoglobin (HHb) and Tissue Oxygenation Index (TOI) in both Vastus Lateralis and Vastus Medialis during rest, exercise, and recovery. Coefficient of Variation (CV), Standard Error Measurement (SEM) and Intraclass Correlation Coefficient (ICC) were used to evaluate the reliability of the data. Significance was set at p < 0.05.
The results indicated that TOI had good to acceptable absolute reliability (CV = 2.7-10.2 %). A good relative relativity for the overall test was found for Vastus Medialis OHb (ICC = 0.851), HHb (ICC = 0.852), and TOI (ICC = 0.864), and Vastus Lateralis OHb (ICC = 0.898), HHb (ICC = 0.899), and TOI (ICC = 0.897).
We conclude that NIRO200Nx is a reliable instrument for measuring muscle oxygen saturation through the TOI parameter in not-to-failure dynamic resistance exercises (1 set of 8 reps against ∼40 % 1 repetition maximum). Tissue oxygenation assessment could be a new way of individualizing exercise through dynamic resistance exercises.
本研究旨在确定一种非便携式近红外光谱仪(NIRO200Nx)在年轻健康男性进行全蹲运动和恢复期间的测试-重测信度。
这项原始研究共招募了 25 名男性参与者(21.8±2.6 岁)。每位参与者完成了 8 次重复测试,负载产生的速度为 1 m·s。该测试进行了两次,两次测试之间有 48 小时的洗脱期。
结果表明,TOI 具有良好到可接受的绝对可靠性(CV=2.7-10.2%)。发现整个测试的 Vastus Medialis OHb(ICC=0.851)、HHb(ICC=0.852)和 TOI(ICC=0.864)、Vastus Lateralis OHb(ICC=0.898)、HHb(ICC=0.899)和 TOI(ICC=0.897)的相对相关性良好。
我们得出结论,NIRO200Nx 是一种可靠的仪器,可通过 TOI 参数测量肌肉氧饱和度,用于非衰竭性动力抗阻运动(1 组 8 次,约 40%的 1 次重复最大值)。组织氧合评估可能是通过动力抗阻运动个性化运动的新方法。