Impact of Nodular Calcifications in the Aortic Annulus and Left Ventricular Outflow Tract on TAVI Outcome with New-Generation Devices.

作者信息

Gorla Riccardo, Oliva Omar A, Poletti Enrico, Finotello Alice, Morganti Simone, Zannoni Jessica, Agnifili Mauro, Barletta Marta, Squillace Mattia, Criscione Enrico, Tusa Maurizio, Brambilla Nedy, Auricchio Ferdinando, Testa Luca, Bedogni Francesco

机构信息

Department of Clinical and Interventional Cardiology, IRCCS Policlinico San Donato, 20097 San Donato Milanese (MI), Italy.

Department of Civil Engineering and Architecture, University of Pavia, 27100 Pavia, Italy.

出版信息

Rev Cardiovasc Med. 2022 Oct 25;23(11):358. doi: 10.31083/j.rcm2311358. eCollection 2022 Nov.

Abstract

BACKGROUND

The impact of nodular calcifications in left ventricular outlow tract (LVOT) and aortic annulus on the procedural outcome of transcatheter aortic valve implantation (TAVI) with new-generation devices is yet to be elucidated. Similarly, computational simulations may provide a novel insight into the biomechanical features of TAVI devices and their interaction with nodular calcifications.

METHODS

This retrospective single-center study included 232 patients submitted to TAVI with Evolut-R (53.4%), Portico (33.6%) and Lotus (13.0%) devices with available preoperative computed tomography (CT) angiography and evidence of nodular calcifications in aortic annulus and/or LVOT. Calcification severity was defined moderate in presence of at least two nodules or one nodule 5 mm. Three virtual simulation models of aortic root presenting a nodular calcification of increasing size were implemented. Stress distribution, stent-root contact area and paravalvular orifice area were computed.

RESULTS

At least moderate calcifications were found in 123 (53.0%) patients, with no sex differences. Among the moderate calcification group, lower device success rate was evident (87.8% vs. 95.4%; = 0.039). Higher rates of moderate paravalvular leak (PVL) (11.4% vs. 3.7%; = 0.028) and vascular complications (9.8% vs. 2.8%; = 0.030) were also observed. Among the Evolut-R group, higher rates of at moderate PVL (12.1%) were observed compared to Portico (3.8%; = 0.045) and Lotus (0.0%; = 0.044) groups. Calcification of both annulus and LVOT (odds ratio [OR] 0.105; = 0.023) were independent predictors of device success. On computational simulations, Portico exhibited homogeneous stress distribution by increasing calfications and overall a larger paravalvular orifice areas compared to Evolut-R and Lotus. Evolut-R showed higher values of average stress than Portico, although with a more dishomogeneous distribution leading to greater paravalvular orifice areas by severe calcifications. Lotus showed overall small paravalvular orifice areas, with no significant increase across the three models.

CONCLUSIONS

At least moderate nodular calcifications in the annulus/LVOT region significantly affected TAVI outcome, as they were independent predictors of device success. Lotus and Portico seemed to perform better than Evolut-R as for device success and moderate PVL. Computational simulations revealed unique biomechanical features of the investigated devices in terms of stent compliance and radial force.

摘要

背景

左心室流出道(LVOT)和主动脉瓣环中的结节状钙化对新一代经导管主动脉瓣植入术(TAVI)手术结果的影响尚待阐明。同样,计算机模拟可能会为TAVI装置的生物力学特征及其与结节状钙化的相互作用提供新的见解。

方法

这项回顾性单中心研究纳入了232例行TAVI手术的患者,使用Evolut-R(53.4%)、Portico(33.6%)和Lotus(13.0%)装置,术前有可用的计算机断层扫描(CT)血管造影,且有主动脉瓣环和/或LVOT结节状钙化的证据。钙化严重程度定义为存在至少两个结节或一个直径≥5mm的结节时为中度。实施了三个主动脉根部虚拟模拟模型,其呈现出大小不断增加的结节状钙化。计算应力分布、支架与根部的接触面积以及瓣周漏口面积。

结果

123例(53.0%)患者发现至少中度钙化,无性别差异。在中度钙化组中,装置成功率明显较低(87.8%对95.4%;P = 0.039)。中度瓣周漏(PVL)发生率(11.4%对3.7%;P = 0.028)和血管并发症发生率(9.8%对2.8%;P = 0.030)也较高。在Evolut-R组中,与Portico组(3.8%;P = 0.045)和Lotus组(0.0%;P = 0.044)相比,中度及以上PVL发生率更高(12.1%)。瓣环和LVOT的钙化(优势比[OR] 0.105;P = 0.023)是装置成功的独立预测因素。在计算机模拟中,与Evolut-R和Lotus相比,Portico随着钙化增加表现出均匀的应力分布,且总体瓣周漏口面积更大。Evolut-R的平均应力值高于Portico,尽管分布更不均匀,严重钙化时导致更大的瓣周漏口面积。Lotus总体瓣周漏口面积较小,在三个模型中无显著增加。

结论

瓣环/LVOT区域至少中度的结节状钙化显著影响TAVI结果,因为它们是装置成功的独立预测因素。在装置成功率和中度及以上PVL方面,Lotus和Portico似乎比Evolut-R表现更好。计算机模拟揭示了所研究装置在支架顺应性和径向力方面独特的生物力学特征。