Suppr超能文献

使用符合 EARL 标准的重建方法,比较长轴向视野与短轴向视野 PET/CT 的临床性能。

Clinical Performance Comparison of a Long Versus a Short Axial Field-of-View PET/CT Using EARL-Compliant Reconstructions.

机构信息

Department of Nuclear Medicine and Molecular Imaging, University of Groningen, University Medical Center Groningen, Hanzeplein 1, 9713 GZ, Groningen, The Netherlands.

Department of Biomedical Photonic Imaging, Faculty of Science and Technology, University of Twente, Enchede, The Netherlands.

出版信息

Mol Imaging Biol. 2024 Oct;26(5):780-789. doi: 10.1007/s11307-024-01939-5. Epub 2024 Aug 2.

Abstract

PURPOSE

To ensure comparable PET/CT image quality between or within centres, clinical inter-system performance comparisons following European Association of Nuclear Medicine Research Ltd. (EARL) guidelines is required. In this work the performance of the long axial field-of-view Biograph Vision Quadra is compared to its predecessor, the short axial field-of-view Biograph Vision.

PROCEDURES

To this aim, patients with suspected tumour lesions received a single weight-based (3 MBq/kg) 2-deoxy-2-[F]fluoro-D-glucose injection and underwent routine clinical ( 15 min) scans on the Vision and 3-min scans on the Quadra in listmode in balanced order. Image quality (IQ), image noise (IN), and tumour demarcation (TD) were assessed visually by four nuclear medicine physicians using a 5-point Likert scale and semiquantitative analysis was performed using standardised uptake values (SUVs). Inter-reader agreement was tested using Wilcoxon's signed rank test and the SUVs were statistically compared using a paired t-test.

RESULTS

Twenty patients (mean age, 60 years ± 8.8 [standard deviation], 16 male) were enrolled. Inter-reader agreement ranged from good to very good for IQ and IN (0.62 ≤ W ≤ 0.81), and fair for TD (0.29 ≤ W ≤ 0.39). Furthermore, a significant difference was found for TD (p = 0.015) between the systems, showing improved TD for the Quadra.

CONCLUSION

This study demonstrates that the Quadra can be used in routine clinical practice with multiple PET/CT systems or in multicentre studies. This system provides comparable diagnostic image quality and semiquantitative accuracy, improved TD, and has the advantage of shorter scan durations.

摘要

目的

为了确保 PET/CT 图像质量在不同中心或同一中心之间具有可比性,需要根据欧洲核医学研究协会(EARL)的指南进行临床系统间性能比较。在这项工作中,比较了长轴向视野的 Biograph Vision Quadra 与短轴向视野的 Biograph Vision 在前瞻性患者中的性能。

方法

为此,疑似肿瘤病变的患者接受了一次基于体重的(3MBq/kg)2-脱氧-2-[F]氟-D-葡萄糖注射,并在 Vision 上进行常规临床(15 分钟)扫描,在 Quadra 上以列表模式以平衡顺序进行 3 分钟扫描。由四位核医学医师使用 5 分制 Likert 量表进行视觉评估图像质量(IQ)、图像噪声(IN)和肿瘤边界(TD),并使用标准化摄取值(SUVs)进行半定量分析。使用 Wilcoxon 符号秩检验测试读者间的一致性,并使用配对 t 检验对 SUVs 进行统计学比较。

结果

共纳入 20 名患者(平均年龄 60 岁±8.8 岁[标准差],16 名男性)。IQ 和 IN 的读者间一致性范围为良好至非常好(0.62≤W≤0.81),TD 的读者间一致性为一般(0.29≤W≤0.39)。此外,系统间 TD 存在显著差异(p=0.015),Quadra 的 TD 改善。

结论

这项研究表明,Quadra 可以在常规临床实践中与多个 PET/CT 系统一起使用,或在多中心研究中使用。该系统提供了可比的诊断图像质量和半定量准确性,改善了 TD,并且具有扫描时间更短的优势。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/50c1/11436434/5edd524e40fb/11307_2024_1939_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验