在具有长轴向视野的Biograph Vision Quadra PET/CT系统上进行EARL合规性测量。

EARL compliance measurements on the biograph vision Quadra PET/CT system with a long axial field of view.

作者信息

Prenosil George A, Hentschel Michael, Weitzel Thilo, Sari Hasan, Shi Kuangyu, Afshar-Oromieh Ali, Rominger Axel

机构信息

Department of Nuclear Medicine, Inselspital Bern, Bern University Hospital, University of Bern, 3010, Bern, Switzerland.

Advanced Clinical Imaging Technology, Siemens Healthcare AG, Lausanne, Switzerland.

出版信息

EJNMMI Phys. 2022 Apr 8;9(1):26. doi: 10.1186/s40658-022-00455-1.

Abstract

BACKGROUND

Our aim was to determine sets of reconstruction parameters for the Biograph Vision Quadra (Siemens Healthineers) PET/CT system that result in quantitative images compliant with the European Association of Nuclear Medicine Research Ltd. (EARL) criteria. Using the Biograph Vision 600 (Siemens Healthineers) PET/CT technology but extending the axial field of view to 106 cm, gives the Vision Quadra currently an around fivefold higher sensitivity over the Vision 600 with otherwise comparable spatial resolution. Therefore, we also investigated how the number of incident positron decays-i.e., exposure-affects EARL compliance. This will allow estimating a minimal acquisition time or a minimal applied dose in clinical scans while retaining data comparability.

METHODS

We measured activity recovery curves on a NEMA IEC body phantom filled with an aqueous F solution and a sphere to background ratio of 10-1 according to the latest EARL guidelines. Reconstructing 3570 image sets with varying OSEM PSF iterations, post-reconstruction Gaussian filter full width at half maximum (FWHM), and varying exposure from 59 kDecays/ml (= 3 s frame duration) to 59.2 MDecays/ml (= 1 h), allowed us to determine sets of parameters to achieve compliance with the current EARL 1 and EARL 2 standards. Recovery coefficients (RCs) were calculated for the metrics RC, RC, and RC, and the respective recovery curves were analyzed for monotonicity. The background's coefficient of variation (COV) was also calculated.

RESULTS

Using 6 iterations, 5 subsets and 7.8 mm Gauss filtering resulted in optimal EARL1 compliance and recovery curve monotonicity in all analyzed frames, except in the 3 s frames. Most robust EARL2 compliance and monotonicity were achieved with 2 iterations, 5 subsets, and 3.6 mm Gauss FWHM in frames with durations between 30 s and 10 min. RC only impeded EARL2 compliance in the 10 s and 3 s frames.

CONCLUSIONS

While EARL1 compliance was robust over most exposure ranges, EARL2 compliance required exposures between 1.2 MDecays/ml to 11.5 MDecays/ml. The Biograph Vision Quadra's high sensitivity makes frames as short as 10 s feasible for comparable quantitative images. Lowering EARL2 RC limits closer to unity would possibly even permit shorter frames.

摘要

背景

我们的目的是确定适用于Biograph Vision Quadra(西门子医疗)PET/CT系统的重建参数集,以生成符合欧洲核医学研究协会(EARL)标准的定量图像。使用Biograph Vision 600(西门子医疗)PET/CT技术,但将轴向视野扩展到106厘米,使得Vision Quadra目前在空间分辨率可比的情况下,灵敏度比Vision 600高约五倍。因此,我们还研究了入射正电子衰变的数量(即曝光量)如何影响EARL合规性。这将有助于在保持数据可比性的同时,估计临床扫描中的最短采集时间或最小应用剂量。

方法

我们根据最新的EARL指南,在一个填充了氟水溶液且球体与背景比值为10:1的NEMA IEC体模上测量了活度恢复曲线。通过重建3570组具有不同OSEM PSF迭代次数、重建后高斯滤波器半高宽(FWHM)以及从59 k衰变数/毫升(=3秒帧持续时间)到59.2 M衰变数/毫升(=1小时)的不同曝光量的图像集,我们得以确定符合当前EARL 1和EARL 2标准的参数集。计算了针对RC、RC和RC指标的恢复系数(RC),并分析了各自恢复曲线的单调性。还计算了背景的变异系数(COV)。

结果

使用6次迭代、5个子集和7.8毫米高斯滤波,在所有分析的帧中(除了3秒的帧),实现了最佳的EARL1合规性和恢复曲线单调性。在持续时间为30秒至10分钟的帧中,使用2次迭代、5个子集和3.6毫米高斯FWHM实现了最稳健可靠的EARL2合规性和单调性。仅在10秒和3秒的帧中,RC妨碍了EARL2合规性。

结论

虽然EARL1合规性在大多数曝光范围内都很稳健可靠,但EARL2合规性要求曝光量在1.2 M衰变数/毫升至11.5 M衰变数/毫升之间。Biograph Vision Quadra的高灵敏度使得对于可比的定量图像,短至10秒的帧成为可能。将EARL2 RC限制降低到更接近1甚至可能允许更短的帧。