手持式超声设备在内科肺部超声检查中的准确性:一项观察性研究。

Accuracy of lung ultrasound performed with handheld ultrasound device in internal medicine: an observational study.

机构信息

Department of Experimental and Clinical Medicine, University of Florence, Careggi Hospital, Florence, Italy.

Department of Health Science, University of Florence, Florence, Italy.

出版信息

J Ultrasound. 2024 Dec;27(4):825-830. doi: 10.1007/s40477-024-00941-5. Epub 2024 Aug 3.

Abstract

AIMS

Lung ultrasound (LUS) is increasingly used in Internal Medicine to complement medical examination, documenting pleural and lung conditions. This study aimed to compare the accuracy of handheld ultrasound device (HHUSD) with high-end ultrasound device (HEUSD) in patients with heart failure or pneumonia, also including the assessment of costs and time-savings.

METHODS

In this observational study 72 patients (aged ≥ 18) admitted to Internal Medicine Unit for heart failure or pneumonia underwent LUS plus evaluation of inferior cava vein (ICV) when indicated, using both HHUSD and HEUSD. Each evaluation, independently performed by 2 different experienced operators, included B-lines number, pleural effusion, lung consolidations, ICV ectasia and its respiratory excursions.

RESULTS

Concordance between HHUSD and HEUSD findings was 79.3% ± 17.7 (mean ± SD) for B-lines, 88.6% for pleural effusion, 82.3% for consolidations and 88.7% and 84.9% for ICV ectasia and its respiratory excursions respectively. BMI didn't significantly influence concordance between the two methods. Moreover, examination time (as mean ± SD) was shorter with HHUSD (8 ± 1.5 min) compared to HEUSD (10 ± 2.5 min).

CONCLUSIONS

HHUSD demonstrated high accuracy in detecting B-lines, pleural effusions, lung consolidations and ICV evaluation when compared to HEUSD. Thus, HHUSD, not only is characterized by accessibility, portability, and easy handling due to its small size, but it also offers advantages in terms of saving costs and time, ultimately contributing to faster patient assessment compared to HEUSD.

摘要

目的

肺部超声(LUS)越来越多地被应用于内科,以补充体格检查,记录胸膜和肺部情况。本研究旨在比较手持式超声设备(HHUSD)和高端超声设备(HEUSD)在心力衰竭或肺炎患者中的准确性,同时评估成本和节省时间。

方法

在这项观察性研究中,72 名(年龄≥18 岁)因心力衰竭或肺炎入住内科病房的患者接受了 LUS 检查,并在需要时使用 HHUSD 和 HEUSD 评估下腔静脉(ICV)。每个评估由 2 名不同经验的操作人员独立进行,包括 B 线数量、胸腔积液、肺部实变、ICV 扩张及其呼吸运动。

结果

HHUSD 和 HEUSD 检查结果的一致性为 B 线 79.3%±17.7(平均值±标准差),胸腔积液 88.6%,实变 82.3%,ICV 扩张及其呼吸运动分别为 88.7%和 84.9%。BMI 对两种方法的一致性没有显著影响。此外,HHUSD 的检查时间(平均值±标准差)为 8±1.5 分钟,明显短于 HEUSD 的 10±2.5 分钟。

结论

与 HEUSD 相比,HHUSD 在检测 B 线、胸腔积液、肺部实变和 ICV 评估方面具有较高的准确性。因此,HHUSD 不仅具有便携性、易操作性和小巧尺寸的特点,而且在成本和时间节省方面具有优势,与 HEUSD 相比,可更快地评估患者。