Tsekas Georgios, Zachiu Cornel, Bol Gijsbert H, van den Dobbelsteen Madelon, Meijers Lieke T C, van Lier Astrid L H M W, de Boer Johannes C J, Raaymakers Bas W
Department of Radiotherapy, University Medical Center Utrecht, Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
Phys Imaging Radiat Oncol. 2024 Jun 6;31:100596. doi: 10.1016/j.phro.2024.100596. eCollection 2024 Jul.
This work investigates the use of a multi-2D cine magnetic resonance imaging-based comprehensive motion monitoring (CMM) system for the assessment of prostate intrafraction 3D drifts. The data of six healthy volunteers were analyzed and the values of a clinically-relevant registration quality factor metric exported by CMM were presented. Additionally, the CMM-derived prostate motion was compared to a 3D-based reference and the 2D-3D tracking agreement was reported. Due to the low quality of SI motion tracking (often 2 mm tracking mismatch between anatomical planes) we conclude that further improvements are desirable prior to clinical introduction of CMM for prostate drift corrections.
本研究探讨了基于多二维电影磁共振成像的综合运动监测(CMM)系统在评估前列腺分次内三维漂移中的应用。分析了六名健康志愿者的数据,并给出了CMM导出的临床相关配准质量因子指标值。此外,将CMM得出的前列腺运动与基于三维的参考进行了比较,并报告了二维-三维跟踪一致性。由于矢状面运动跟踪质量较低(解剖平面之间的跟踪失配通常为2毫米),我们得出结论,在将CMM临床应用于前列腺漂移校正之前,需要进一步改进。