Kawahara Daisuke
Department of Radiation Oncology, Institute of Biomedical and Health Sciences, Hiroshima University, Hiroshima, Japan.
Rep Pract Oncol Radiother. 2024 Jul 22;29(3):271-279. doi: 10.5603/rpor.101098. eCollection 2024.
The objective was to enhance the biological compensation factor related to irradiation interruption in a short time (short irradiation interruption) in hypoxic tumors using a refined microdosimetric kinetic model (MKM) for photon radiation therapy.
The biological dose differences were calculated for CHO-K1 cells exposed to a photon beam, considering interruptions of (τ) of 0-120 min and pO at oxygen levels of 0.075-160 mm Hg. The interrupted dose fraction (IDF) was defined as the percentage ratio of the dose delivered before short irradiation interruption to the total dose, which ranged from 10-90%. The compensated dose was calculated based on an IDF of 10-90% for a dose of 2-8 Gy and oxygen levels of 0.075-160 mm Hg.
The Δ with and without short irradiation interruption was more pronounced with a higher dose and increased pO. It exceeded 3% between IDF of 50% and either 10% or 90% and occurred more than τ = 50 min at 0.075 mm Hg, τ = 20 min at 3 mm Hg, τ = 20 min at 8 mm Hg, τ = 20 min at 15 mm Hg, τ = 20 min at 38 mm Hg, and τ = 20 min at 160 mm Hg. The dose compensation factor was greater at higher IDF rates.
The biological dose decreased with longer interruption times and higher oxygen concentrations. The improved model can compensate for the biological doses at various oxygen concentrations.
The current study improved the dose compensation method for the decrease in the biological effect owing to short irradiation interruption by considering the oxygen concentration.
目的是使用用于光子放射治疗的精细微剂量动力学模型(MKM),提高缺氧肿瘤在短时间内(短照射中断)与照射中断相关的生物补偿因子。
计算暴露于光子束的CHO-K1细胞的生物剂量差异,考虑0至120分钟的中断时间(τ)以及0.075至160毫米汞柱氧水平下的pO。中断剂量分数(IDF)定义为短照射中断前给予的剂量与总剂量的百分比,范围为10%至90%。基于2至8戈瑞剂量、0.075至160毫米汞柱氧水平下10%至90%的IDF计算补偿剂量。
有和没有短照射中断时的Δ在较高剂量和升高的pO下更明显。在IDF为50%与10%或90%之间时超过3%,并且在0.075毫米汞柱时τ>50分钟、3毫米汞柱时τ>20分钟、8毫米汞柱时τ>20分钟、15毫米汞柱时τ>20分钟、38毫米汞柱时τ>20分钟以及160毫米汞柱时τ>20分钟时出现。剂量补偿因子在较高的IDF率时更大。
生物剂量随着中断时间延长和氧浓度升高而降低。改进后的模型可以补偿不同氧浓度下的生物剂量。
本研究通过考虑氧浓度改进了因短照射中断导致生物效应降低的剂量补偿方法。