Institute for Integrated Sports Medicine, Keio University School of Medicine, 35 Shinanomachi, Shinjuku-ku, Tokyo, 160-8582, Japan.
Department of Cardiology, Keio University School of Medicine, Tokyo, Japan.
Sci Rep. 2024 Aug 16;14(1):18985. doi: 10.1038/s41598-024-70001-9.
A simple method for determining the anaerobic threshold in patients with heart failure (HF) is needed. This prospective clinical trial (LacS-001) aimed to investigate the safety of a sweat lactate-monitoring sensor and the correlation between lactate threshold in sweat (sLT) and ventilatory threshold (VT). To this end, we recruited 50 patients with HF and New York Heart Association functional classification I-II (mean age: 63.5 years, interquartile range: 58.0-72.0). Incremental exercise tests were conducted while monitoring sweat lactate levels using our sensor. sLT was defined as the first steep increase in lactate levels from baseline. Primary outcome measures were a correlation coefficient of ≥ 0.6 between sLT and VT, similarities as assessed by the Bland-Altman analysis, and standard deviation of the difference within 15 W. A correlation coefficient of 0.651 (95% confidence interval, 0.391-0.815) was achieved in 32/50 cases. The difference between sLT and VT was -4.9 ± 15.0 W. No comparative error was noted in the Bland-Altman plot. No device-related adverse events were reported among the registered patients. Our sweat lactate sensor is safe and accurate for detecting VT in patients with HF in clinical settings, thereby offering valuable additional information for treatment.
需要一种用于确定心力衰竭(HF)患者无氧阈值的简单方法。这项前瞻性临床试验(LacS-001)旨在研究汗液乳酸监测传感器的安全性,以及汗液乳酸阈值(sLT)与通气阈值(VT)之间的相关性。为此,我们招募了 50 名 HF 患者和纽约心脏协会功能分类 I-II 级(平均年龄:63.5 岁,四分位距:58.0-72.0)。使用我们的传感器监测汗液中的乳酸水平进行递增运动测试。sLT 定义为乳酸水平从基线首次急剧升高。主要观察指标为 sLT 和 VT 之间的相关系数≥0.6, Bland-Altman 分析评估的相似性,以及 15 W 内差异的标准差。在 32/50 例中,获得了 0.651(95%置信区间,0.391-0.815)的相关系数。sLT 和 VT 之间的差异为-4.9±15.0 W。Bland-Altman 图中没有观察到比较误差。在登记的患者中没有报告与设备相关的不良事件。我们的汗液乳酸传感器在临床环境中用于检测 HF 患者的 VT 是安全且准确的,从而为治疗提供了有价值的附加信息。