Department of Endodontics, Faculty of Dentistry, Akdeniz University, Antalya, Turkey, 07070.
Health Sciences Institute, Department of Medical Biotechnology, Akdeniz University, Antalya, Turkey, 07070.
BMC Oral Health. 2024 Aug 16;24(1):953. doi: 10.1186/s12903-024-04713-9.
To evaluate the accuracy of the electronic apex locators (EALs), and Cone-Beam Computed Tomography (CBCT) scanning, both in working length (WL) determination and in the detection of root canal perforations in retreatment cases.
Sixty human mandibular premolars were selected. After crown removal partially and canal access, root canals were instrumented and irrigated. The obturation process utilized gutta-percha and sealer with warm vertical compaction. Two groups were distinguished: one without perforation (Group 1) and the other with an apical third perforation (Group 2). Retreatment included filling removal, apical preparation, and irrigation. Actual working lengths (AWL) were determined using a stereomicroscope. CBCT images were used to measure CBCT working length (CWL), with adjustments for optimal views. Propex II and Dentaport ZX were used to measure electronic working length (EWL). Differences between EWL and AWL, as well as CWL, were analyzed to gauge accuracy. Data underwent Two-way ANOVA analysis. Measurements within ± 0.5 and ± 1 mm tolerance ranges were deemed successful for each device, followed by applying the Pearson Chi-square test.
The study reveals no significant inter-group variations in device performance (p > .05). Dentaport ZX missed detecting perforation in two Group 2 (apical perforation) cases. For ± 1 mm tolerance, Propex II displayed the highest success in Group 2 (apical perforation).
This study demonstrates the comparable performance of Propex II, Dentaport ZX, and CBCT in endodontic retreatment, providing insights into diagnostic reliability.
评估电子根尖定位仪(EAL)和锥形束 CT(CBCT)扫描在确定工作长度(WL)和检测再治疗病例根管穿孔方面的准确性。
选择 60 个人类下颌前磨牙。去除部分冠部并进行根管通路预备后,对根管进行清理和冲洗。使用热垂直加压技术,用牙胶和密封剂进行根管充填。将样本分为两组:一组无穿孔(组 1),另一组有根尖三分之一穿孔(组 2)。再治疗包括去除填充物、根尖预备和冲洗。使用立体显微镜确定实际工作长度(AWL)。使用 CBCT 图像测量 CBCT 工作长度(CWL),并进行最佳视图调整。使用 Propex II 和 Dentaport ZX 测量电子工作长度(EWL)。分析 EWL 与 AWL 以及 CWL 之间的差异,以评估准确性。数据采用双向方差分析。对于每个设备,如果测量值在±0.5 和±1 毫米的容差范围内,则认为测量成功,然后应用 Pearson 卡方检验。
研究结果显示,各设备之间在组间差异上无统计学意义(p>.05)。在两组中,Dentaport ZX 各漏诊了 2 例根尖穿孔。在±1 毫米容差范围内,Propex II 在组 2(根尖穿孔)中显示出最高的成功率。
本研究表明 Propex II、Dentaport ZX 和 CBCT 在根管再治疗中的性能相当,为诊断可靠性提供了参考。