临床条件下电子根尖定位仪Dentaport ZX和Ray Pex 4准确性的评估。

Evaluation of the accuracy of electronic apex locators Dentaport ZX and Ray Pex 4 under clinical conditions.

作者信息

Stavrianos Christos, Vladimirov Stoyan B, Vangelov Ljubomir S, Papadopoulos C, Bouzala A

机构信息

Aristotle University of Thessaloniki, School of Dentistry, Department of Endodontology, Thessaloniki, Greece.

出版信息

Folia Med (Plovdiv). 2007;49(3-4):75-9.

Abstract

AIM

The aim of the present study was to compare the accuracy of working endodontic length determination by Dentaport ZX and Ray Pex 4 under clinical conditions.

MATERIAL AND METHODS

Clinical comparative evaluation of the accuracy of two electronic apex locators, Dentaport ZX and Ray-Pex 4, was carried out in 80 single-rooted teeth with vital pulp tissue planned for extraction as part of the dental treatment and after patient consent had been obtained. The normal procedure was followed, which included anesthesia, a standard endodontic access cavity, pulp removal and irrigation of the root canal with 5.25% NaOCl. The working length was estimated with Dentaport ZX for half of the selected teeth and with Ray-Pex 4 for the other half. K-files #15 were cemented at the measured working length and the teeth were extracted. The apical foramen was examined with a stereomicroscope under x15 magnification. The micrometric system of the microscope allowed precise measurement of distances.

RESULTS

The comparison of the estimated working length with the actual length revealed that Dentaport ZX localized the apical foramen in 95% of the cases and the Ray Pex 4 in 92.5% with a clinically acceptable accuracy of +/- 0.5 mm. Dentaport ZX detected the foramen in 97.5% and Ray Pex 4 in 95% of the cases with an accuracy of +/- 1 mm. There was no statistically significant difference between the accuracy of the two electronic apex locators in locating the apical foramen (p < 0.05).

摘要

目的

本研究的目的是在临床条件下比较Dentaport ZX和Ray Pex 4测定根管工作长度的准确性。

材料与方法

在80颗有活力牙髓组织且计划拔除作为牙科治疗一部分的单根牙中,经患者同意后,对两种电子根尖定位仪Dentaport ZX和Ray-Pex 4的准确性进行临床对比评估。遵循常规程序,包括麻醉、标准的根管入口洞形、牙髓摘除以及用5.25%次氯酸钠冲洗根管。对所选牙齿的一半用Dentaport ZX估计工作长度,另一半用Ray-Pex 4估计工作长度。将15号K锉粘结在测量的工作长度处,然后拔除牙齿。在15倍放大倍数下用体视显微镜检查根尖孔。显微镜的测微系统能够精确测量距离。

结果

估计工作长度与实际长度的比较显示,Dentaport ZX在95%的病例中定位到根尖孔,Ray Pex 4在92.5%的病例中定位到根尖孔,临床可接受的精度为±0.5毫米。Dentaport ZX在97.5%的病例中检测到根尖孔,Ray Pex 4在95%的病例中检测到根尖孔,精度为±1毫米。两种电子根尖定位仪在定位根尖孔的准确性上无统计学显著差异(p<0.05)。