Her Jiwoong, Roh Justin, Keys Deborah A
Department of Veterinary Clinical Sciences, College of Veterinary Medicine, The Ohio State University, Columbus, OH, USA.
Department of Clinical Sciences, College of Veterinary Medicine, North Carolina State University, Raleigh, NC, USA.
Vet Med (Auckl). 2024 Aug 16;15:197-203. doi: 10.2147/VMRR.S465945. eCollection 2024.
To assess the agreement between pulse co-oximeter and blood co-oximeter in measuring methemoglobin (MetHb) and carboxyhemoglobin (COHb) in healthy, awake dogs.
Forty-five healthy dogs were enrolled in the study. Pulse co-oximetry was performed using the Masimo Radical 7 Pulse Co-Oximeter with a Rainbow adhesive sensor. Simultaneously, venous blood samples were collected, and MetHb and COHb were immediately measured using a Stat Profile Prime Plus VET Critical Care Analyzer. Paired measurements of MetHb and COHb were evaluated via Spearman correlation, intra-class correlation (ICC), and Bland-Altman plots to evaluate the degree of agreement between the pulse co-oximeter and the blood co-oximeter.
A total of 45 paired MetHb and COHb measurements were collected. There was a weak correlation between the pulse co-oximeter and the blood co-oximeter readings. The correlation coefficients for MetHb and COHb were 0.0 (95% CI, -0.3 to 0.3) and 0.03 (95% CI, -0.27 to 0.32), respectively. The ICC indicated poor agreement between the pulse and blood co-oximeter for MetHb (ICC = 0.00, 95% CI: -0.12 to 0.15) and COHb (ICC = 0.03, 95% CI: -0.27 to 0.33). Bland-Altman plots revealed low mean bias but wide limits of agreement, indicating that the pulse co-oximeter overestimated MetHb by on average of 0.7% (P < 0.0001) (95% LoA: -0.5 to 2.0) and COHb by on average 0.2% (P = 0.59) (95% LoA: -4.6 to 5.0).
Obtaining MetHb and COHb measurements with the Masimo Radical 7 Pulse Co-Oximeter is straightforward in healthy, awake dogs. However, the device does not provide accurate measurements compared to the blood co-oximeter, specifically in the range of MetHb and COHb in healthy dogs, based on the wide LoA.
评估脉搏血氧仪与血气分析仪在测量健康清醒犬高铁血红蛋白(MetHb)和碳氧血红蛋白(COHb)时的一致性。
45只健康犬纳入本研究。使用带有Rainbow粘性传感器的Masimo Radical 7脉搏血氧仪进行脉搏血氧测定。同时采集静脉血样,立即使用Stat Profile Prime Plus VET重症监护分析仪测量MetHb和COHb。通过Spearman相关性、组内相关性(ICC)和Bland-Altman图对MetHb和COHb的配对测量结果进行评估,以评价脉搏血氧仪与血气分析仪之间的一致性程度。
共收集到45对MetHb和COHb测量值。脉搏血氧仪与血气分析仪读数之间存在弱相关性。MetHb和COHb的相关系数分别为0.0(95%CI,-0.3至0.3)和0.03(95%CI,-0.27至0.32)。ICC表明,脉搏血氧仪与血气分析仪在测量MetHb(ICC = 0.00,95%CI:-0.12至0.15)和COHb(ICC = 0.03,95%CI:-0.27至0.33)方面一致性较差。Bland-Altman图显示平均偏差较低,但一致性界限较宽,表明脉搏血氧仪高估MetHb平均0.7%(P < 0.0001)(95%一致性界限:-0.5至2.0),高估COHb平均0.2%(P = 0.59)(95%一致性界限:-4.6至5.0)。
在健康清醒犬中,使用Masimo Radical 7脉搏血氧仪获取MetHb和COHb测量值操作简便。然而,基于较宽的一致性界限,与血气分析仪相比,该设备不能提供准确的测量值,特别是在健康犬的MetHb和COHb范围内。