Department of Clinical Biochemistry, Aarhus University Hospital, Aarhus, Denmark.
Department of Clinical Medicine, Aarhus University, Aarhus, Denmark.
Scand J Clin Lab Invest. 2024 Oct;84(6):421-424. doi: 10.1080/00365513.2024.2394981. Epub 2024 Aug 24.
It is common practice in laboratories to store biological samples in ultra-low temperature (ULT) freezers. There is growing interest in raising the temperature of ULT freezers in order to save energy and reduce expenses, as energy conservation becomes increasingly important and sustainable laboratory practices gain popularity. In our laboratory, plasma samples are stored for three months for diagnostic purposes. We therefore took the opportunity to investigate the effect of two different storage temperatures (-70 °C vs -80 °C), on activated partial thromboplastin time (APTT), factor VIII (FVIII), international normalized ratio (INR) and factor VII (FVII) measurements on paired plasma samples collected from 26 individuals after three months of storage. Automated coagulation analysers CS-5100 and ACL TOP were used to perform the tests. We found no consistent difference between the two storage temperatures for any of the four coagulation parameters (all -values > 0.05). We conclude that the temperature of ULT freezers used to store plasma samples for APTT, FVIII, INR, and FVII measurements can be safely increased from -80 to -70 °C without affecting the stability of the samples.
在实验室中,将生物样本储存在超低温(ULT)冰箱中是一种常见的做法。为了节约能源和降低成本,提高 ULT 冰箱的温度越来越受到关注,因为节能变得越来越重要,可持续的实验室实践也越来越受欢迎。在我们的实验室中,为了诊断目的,血浆样本储存三个月。因此,我们有机会研究了两种不同的储存温度(-70°C 与-80°C)对 26 名个体储存三个月后的配对血浆样本中激活部分凝血活酶时间(APTT)、VIII 因子(FVIII)、国际标准化比值(INR)和 VII 因子(FVII)测量值的影响。我们使用 CS-5100 和 ACL TOP 自动化凝血分析仪进行了测试。我们发现,对于四个凝血参数中的任何一个,两种储存温度之间没有一致的差异(所有 - 值 > 0.05)。我们得出结论,用于储存 APTT、FVIII、INR 和 FVII 测量的血浆样本的 ULT 冰箱的温度可以安全地从-80°C 升高到-70°C,而不会影响样本的稳定性。