Suppr超能文献

关于 B0 匀场算法对单体磁共振波谱的影响。

On the impact of B0 shimming algorithms on single-voxel MR spectroscopy.

机构信息

NeuroPoly Lab, Institute of Biomedical Engineering, Polytechnique Montréal, Montréal, Quebec, Canada.

Centre de recherche du CHU Sainte-Justine, Montréal, Quebec, Canada.

出版信息

Magn Reson Med. 2025 Jan;93(1):42-50. doi: 10.1002/mrm.30257. Epub 2024 Aug 26.

Abstract

PURPOSE

To assess the impact of different B0 shimming algorithms on MRS.

METHODS

B0 field maps and single-voxel MR spectroscopy were acquired in the prefrontal cortex of five volunteers at 3 T using five different B0 shimming approaches. B0 shimming was achieved using Siemens' proprietary shim algorithm, in addition to the Pseudo-Inverse (PI), Quadratic Programming (QuadProg), Least Squares (LSq), and Gradient optimization (Grad) algorithms. The standard deviation of the shimmed B0 field, as well as the SNR and FWHM of the measured metabolites, was used to evaluate the performance of each B0 shimming algorithm.

RESULTS

Compared to Siemens's shim, significant reductions (p < 0.01) in the standard deviation of the B0 field distribution within the MRS voxel were observed for the PI, QuadProg, and Grad algorithms (3.8 Hz, 7.3 Hz, and 3.9 Hz respectively, compared to 11.5 Hz for Siemens), but not for the LSq (12.9 Hz) algorithm. Moreover, significantly increased SNR and reduced FWHM for the N-acetylaspartate metabolite were consistent with the improvement in B0 homogeneity for the aforementioned shimming algorithms.

CONCLUSION

Here, we demonstrate that the choice of B0 shimming algorithm can have a significant impact on the quality of MR spectra and that significant improvements in spectrum quality could be achieved by using alternatives to the default vendor approach.

摘要

目的

评估不同 B0 匀场算法对 MRS 的影响。

方法

在 3T 下,使用五种不同的 B0 匀场方法,在五名志愿者的前额叶皮层采集 B0 磁场图和单体素磁共振波谱。使用西门子专有的匀场算法以及伪逆(PI)、二次规划(QuadProg)、最小二乘(LSq)和梯度优化(Grad)算法实现 B0 匀场。评估每个 B0 匀场算法的性能,使用匀场后 B0 磁场的标准差、测量代谢物的 SNR 和 FWHM。

结果

与西门子的匀场算法相比,PI、QuadProg 和 Grad 算法的 MRS 体素内 B0 磁场分布的标准差显著降低(p<0.01)(分别为 3.8Hz、7.3Hz 和 3.9Hz,而西门子为 11.5Hz),但 LSq(12.9Hz)算法则没有。此外,N-乙酰天门冬氨酸代谢物的 SNR 显著增加和 FWHM 减小与上述匀场算法的 B0 均匀性改善一致。

结论

本研究表明,B0 匀场算法的选择会对磁共振谱的质量产生重大影响,并且通过使用替代默认供应商方法,可以显著提高谱的质量。