Tanaka Mami, Umemoto Yasutaka, Ohashi Wataru, Watanabe Hideto, Nagata Ayako, Furukawa Hiroshi
From the Department of Plastic and Reconstructive Surgery, Aichi Medical University, Nagakute, Japan.
Clinical Research Center, Aichi Medical University, Nagakute, Japan.
Plast Reconstr Surg Glob Open. 2024 Aug 26;12(8):e6096. doi: 10.1097/GOX.0000000000006096. eCollection 2024 Aug.
The deep inferior epigastric perforator (DIEP) flap is a useful tool for breast reconstruction and tends to be transferred into the breast envelope as the buried flap from an aesthetic point of view. However, it is difficult to monitor the blood flow in the buried DIEP flap after reconstructive microsurgery. Near-infrared spectroscopy devices have recently been used for monitoring the blood flow of various organs. NIRO200NX (Hamamatsu Photonics) continuously measures the tissue oxygen index (TOI) and quickly reflects changes in flap blood flow. In this study, we investigated whether and how much the NIRO200NX applies to monitoring the blood flow of the buried flap.
We included 156 patients who underwent breast reconstruction using a DIEP flap from October 2013 to May 2022, comprising 57 exposed and 99 buried-type DIEP flap cases. We measured TOI using NIRO200NX, in combination with conventional evaluation methods, including color check, pinprick test, and Doppler sound.
A criterion of TOI 50 gave the best evaluations. All the 57 exposed-type flap cases showed no false evaluations, and NIRO200NX performed precise judgment. In 99 buried-type flap cases, NIRO200NX correctly evaluated 96 cases. For those buried-type cases, we found only two false-positive and one false-negative case. The misjudgments by NIRO200NX were likely caused by hematoma.
We propose NIRO200NX as a reliable device for monitoring the blood flow of the DIEP flap and predicting the outcomes of breast reconstruction by the DIEP flap transfer.
腹壁下深动脉穿支(DIEP)皮瓣是乳房重建的一种有用工具,从美学角度来看,它倾向于作为埋藏皮瓣转移至乳房包膜内。然而,在重建显微外科手术后,很难监测埋藏的DIEP皮瓣中的血流情况。近年来,近红外光谱设备已被用于监测各种器官的血流。NIRO200NX(滨松光子学公司)可连续测量组织氧合指数(TOI),并能快速反映皮瓣血流的变化。在本研究中,我们调查了NIRO200NX是否适用于监测埋藏皮瓣的血流以及适用程度。
我们纳入了2013年10月至2022年5月期间接受DIEP皮瓣乳房重建的156例患者,其中包括57例外露型和99例埋藏型DIEP皮瓣病例。我们使用NIRO200NX测量TOI,并结合包括颜色检查、针刺试验和多普勒声音在内的传统评估方法。
TOI为50的标准给出了最佳评估结果。所有57例外露型皮瓣病例均未出现错误评估,NIRO200NX做出了准确判断。在99例埋藏型皮瓣病例中,NIRO200NX正确评估了96例。对于那些埋藏型病例,我们仅发现2例假阳性和1例假阴性病例。NIRO200NX的误判可能是由血肿引起的。
我们建议将NIRO200NX作为一种可靠的设备,用于监测DIEP皮瓣的血流,并预测通过DIEP皮瓣转移进行乳房重建的结果。