Florkowski Christopher M, Buchan Vanessa, Li Bobby V, Taylor Felicity, Phan Minh, Than Martin, Pickering John W
63588 Canterbury Health Laboratories , Christchurch Hospital, Christchurch, New Zealand.
Department of Laboratory Medicine, Tan Tock Seng Hospital, Singapore, Singapore.
Clin Chem Lab Med. 2024 Aug 28. doi: 10.1515/cclm-2024-0312.
The Siemens Point-of-Care Testing (POC) Atellica VTLi high-sensitivity troponin I (hsTnI) device has been previously validated. Verification independently provides evidence that an analytical procedure fulfils concordance with laboratory assays, imprecision, and hemolysis interference requirements.
Five whole blood samples spanning the measuring interval were analysed 20 times in succession. Hemolysis interference was assessed at three troponin concentrations by spiking five hemolysate concentrations to plasma to achieve free hemoglobin concentrations 35-1,000 mg/dL. Concordance between whole blood (VTLi) and plasma on laboratory analysers (Beckman, Roche, Siemens) was assessed by Pearson correlation and kappa statistics at the (LOQ) and upper reference limit (URL). This was repeated for frozen plasma samples.
Coefficients of variation for whole blood were <10 % for whole blood troponin concentrations of 9.2 and 15.9 ng/L, thus below the URL. Hemolysis positively interfered; at 250 mg/dL affecting the low troponin sample (+3 ng/L; +60 %) and high troponin sample (+37 ng/L; +24 %). Correlation coefficients were 0.98, 0.90 and 0.97 between VTLi and Beckman, Roche and Siemens assays respectively. Corresponding kappa statistics were 0.80, 0.73 and 0.84 at the LOQ and 0.70, 0.44 and 0.67 at the URL.
Concordances between VTLi and laboratory assays were at least non-inferior to those between laboratory assays. Imprecision met manufacturer claims and was consistent with a high sensitivity assay. There is potential for hemolysis interference, highlighting the need for quality samples. The results support performance characteristics previously reported in validation studies, and the device offers acceptable performance for use within intended medical settings.
西门子即时检验(POC)Atellica VTLi高敏肌钙蛋白I(hsTnI)检测设备先前已通过验证。验证独立提供证据,证明一种分析程序符合实验室检测的一致性、不精密度和溶血干扰要求。
对跨越测量区间的5份全血样本连续进行20次分析。通过向血浆中加入5种溶血产物浓度以达到35 - 1000mg/dL的游离血红蛋白浓度,在3种肌钙蛋白浓度下评估溶血干扰。通过Pearson相关性和kappa统计量在检测限(LOQ)和参考上限(URL)评估全血(VTLi)与实验室分析仪(贝克曼、罗氏、西门子)上的血浆之间的一致性。对冷冻血浆样本重复此操作。
全血肌钙蛋白浓度为9.2和15.9ng/L时,全血变异系数<10%,因此低于URL。溶血产生正向干扰;在250mg/dL时,影响低肌钙蛋白样本(增加3ng/L;增加60%)和高肌钙蛋白样本(增加37ng/L;增加24%)。VTLi与贝克曼、罗氏和西门子检测之间的相关系数分别为0.98、0.90和0.97。在LOQ时相应的kappa统计量分别为0.80、0.73和0.84,在URL时分别为0.70、0.44和0.67。
VTLi与实验室检测之间的一致性至少不劣于实验室检测之间的一致性。不精密度符合制造商声明,且与高灵敏度检测一致。存在溶血干扰的可能性,突出了对质量样本的需求。结果支持先前在验证研究中报告的性能特征,并且该设备在预期医疗环境中使用具有可接受的性能。