基于智能手机系统和金标准足底压力测试系统的步态线性和非线性变异性指标的比较

Validation of Linear and Nonlinear Gait Variability Measures Derived From a Smartphone System Compared to a Gold-Standard Footswitch System During Overground Walking.

机构信息

School of Kinesiology and Health Science, York University, Toronto, ON, Canada.

出版信息

J Appl Biomech. 2024 Sep 2;40(5):437-443. doi: 10.1123/jab.2022-0293. Print 2024 Oct 1.

Abstract

Smartphones, with embedded accelerometers, may be a viable method to monitor gait variability in the free-living environment. However, measurements estimated using smartphones must first be compared to known quantities to ensure validity. This study assessed the validity and reliability of smartphone-derived gait measures compared to a gold-standard footswitch system during overground walking. Seventeen adults completed three 8-minute overground walking trials during 3 separate visits. The stride time series was calculated as the time difference between consecutive right heel contact events within the footswitch and smartphone-accelerometry signals. Linear (average stride time, stride time standard deviation, and stride time coefficient of variation) and nonlinear (fractal scaling index, approximate entropy, and sample entropy) measures were calculated for each stride time series. Bland-Altman plots with 95% limits of agreement assessed agreement between systems. Intraclass correlation coefficients assessed reliability across visits. Bland-Altman plots revealed acceptable limits of agreement for all measures. Intraclass correlation coefficients revealed good-to-excellent reliability for both systems, except for fractal scaling index, which was moderate. The smartphone system is a valid method and performs similarly to gold-standard research equipment. These findings suggest the development and implementation of an inexpensive, easy-to-use, and ubiquitous telehealth instrument that may replace traditional laboratory equipment for use in the free-living environment.

摘要

智能手机配备嵌入式加速度计,可能是一种在自由活动环境中监测步态变化的可行方法。然而,使用智能手机估计的测量值必须首先与已知数量进行比较,以确保其有效性。本研究评估了智能手机衍生的步态测量值与地面行走时黄金标准脚踏开关系统相比的有效性和可靠性。17 名成年人在 3 次单独的访问中完成了 3 次 8 分钟的地面行走试验。步长时间序列是通过计算脚踏开关和智能手机加速度计信号中连续右足跟触事件之间的时间差来计算的。为每个步长时间序列计算了线性(平均步长时间、步长时间标准差和步长时间变异系数)和非线性(分形标度指数、近似熵和样本熵)指标。95%一致性界限的 Bland-Altman 图评估了系统之间的一致性。组内相关系数评估了各次访问的可靠性。 Bland-Altman 图显示所有指标的一致性界限都可以接受。组内相关系数显示两种系统均具有良好到极好的可靠性,除了分形标度指数,其可靠性为中等。智能手机系统是一种有效的方法,其性能与黄金标准研究设备相似。这些发现表明,开发和实施一种经济实惠、易于使用且无处不在的远程医疗仪器是可行的,它可以替代传统的实验室设备,用于自由活动环境。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验