[Aortic and mitral valve bioprostheses. Normal and pathological M mode echocardiographic aspects].

作者信息

Marino J P, Issad M S, Fernandez F, Tarzi E, Baragan J, Gerbaux A, Gay J

出版信息

Arch Mal Coeur Vaiss. 1985 Apr;78(4):580-91.

PMID:3923985
Abstract

The M mode echocardiographic recordings of 52 normal mitral bioprostheses (NMB), 7 pathological mitral bioprostheses (PMB), 30 normal aortic bioprostheses (NAB) and 10 pathological aortic bioprostheses (PAB) were reviewed. In normal bioprostheses a significant correlation was observed between the echocardiographic and the "specified" diameters, the diastolic and systolic slopes and the amplitude of anterior motion of the support. In NMB, the end-systolic diameter of the left ventricular outflow tract depended on the "specified" diameter of the bioprosthesis. Paradoxical septal motion was observed in 78 p. 100 of cases. In PMB, the velocity of anterior leaflet opening was significantly increased (p less than 0.001). The end-diastolic internal left ventricular dimension was also increased (p less than 0.01). A significant correlation was found between left ventricular fractional shortening and maximal leaflet separation (p less than 0.05). Normal septal motion was more common (p less than 0.05). In 5 cases of prosthetic valve dysfunction with mitral regurgitation the maximal leaflet separation was greater than normal (p less than 0.001), the diastolic slope of the support was increased (p less than 0.05) and diastolic vibrations of thickened irregular leaflets were observed. Systolo-diastolic vibrations with chaotic leaflet motion were characteristic of cusp tear and/or eversion. Stratified echos behind a support with reduced leaflet excursion was observed in one case of partial thrombosis: a thickened systolic echo with reduced diastolic excursion was observed in a case of degenerative stenosis. The review of 10 PAB showed a reduced amplitude of systolic excursion of the anterior support in cases of aortic regurgitation (p less than 0.05). Systolic vibrations of the cusp were not specific and were observed in normal cases. In severe valvular regurgitation mitral and/or septal diastolic fluttering was observed. Systolic excursion of the cusps was reduced in cases of relative stenosis due to an inappropriately small sized bioprosthesis. Thickening of the diastolic cusp echos was observed in cases of degenerative stenosis. Ventricular dilatation and reduced septal and free wall motion were dysfunction.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验