[Aortic and mitral valve bioprostheses. Normal and pathological M mode echocardiographic aspects].

作者信息

Marino J P, Issad M S, Fernandez F, Tarzi E, Baragan J, Gerbaux A, Gay J

出版信息

Arch Mal Coeur Vaiss. 1985 Apr;78(4):580-91.

PMID:3923985
Abstract

The M mode echocardiographic recordings of 52 normal mitral bioprostheses (NMB), 7 pathological mitral bioprostheses (PMB), 30 normal aortic bioprostheses (NAB) and 10 pathological aortic bioprostheses (PAB) were reviewed. In normal bioprostheses a significant correlation was observed between the echocardiographic and the "specified" diameters, the diastolic and systolic slopes and the amplitude of anterior motion of the support. In NMB, the end-systolic diameter of the left ventricular outflow tract depended on the "specified" diameter of the bioprosthesis. Paradoxical septal motion was observed in 78 p. 100 of cases. In PMB, the velocity of anterior leaflet opening was significantly increased (p less than 0.001). The end-diastolic internal left ventricular dimension was also increased (p less than 0.01). A significant correlation was found between left ventricular fractional shortening and maximal leaflet separation (p less than 0.05). Normal septal motion was more common (p less than 0.05). In 5 cases of prosthetic valve dysfunction with mitral regurgitation the maximal leaflet separation was greater than normal (p less than 0.001), the diastolic slope of the support was increased (p less than 0.05) and diastolic vibrations of thickened irregular leaflets were observed. Systolo-diastolic vibrations with chaotic leaflet motion were characteristic of cusp tear and/or eversion. Stratified echos behind a support with reduced leaflet excursion was observed in one case of partial thrombosis: a thickened systolic echo with reduced diastolic excursion was observed in a case of degenerative stenosis. The review of 10 PAB showed a reduced amplitude of systolic excursion of the anterior support in cases of aortic regurgitation (p less than 0.05). Systolic vibrations of the cusp were not specific and were observed in normal cases. In severe valvular regurgitation mitral and/or septal diastolic fluttering was observed. Systolic excursion of the cusps was reduced in cases of relative stenosis due to an inappropriately small sized bioprosthesis. Thickening of the diastolic cusp echos was observed in cases of degenerative stenosis. Ventricular dilatation and reduced septal and free wall motion were dysfunction.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验