J Refract Surg. 2024 Sep;40(9):e645-e653. doi: 10.3928/1081597X-20240718-02. Epub 2024 Sep 1.
To evaluate the intrasession repeatability of wavefront aberrations obtained by a combined adaptive optics visual simulator and Hartman-Shack aberrometer in pseudophakic eyes with and without previous corneal refractive surgery.
Three consecutive measurements were performed in one eye of each individual. Total ocular aberrations were recorded up to the 5th Zernike order for a 4.5-mm pupil. Repeatability was assessed by calculating the within-subject standard deviation (S), the repeatability limit (R), and the intraclass correlation coefficient (ICC). Vector analysis was performed to assess astigmatism variability between scans.
The study enrolled 32 normal individuals and 24 individuals with a history of refractive surgery. In normal and eyes that had previous refractive surgery, respectively, the S values were 0.155 and 0.176 diopters (D) for sphere and 0.184 and 0.265 D for cylinder. The S values for all 3rd order terms ranged from 0.037 to 0.047 µm in normal eyes and 0.044 to 0.063 µm in eyes that had previous refractive surgery. The S for primary spherical aberration was 0.020 µm in normal eyes and 0.026 µm in eyes that had previous refractive surgery. ICC values for measurements of astigmatism yielded larger variability (ICC = 0.751 and 0.879). However, both groups demonstrated excellent repeatability (ICC > 0.9) for root mean square higher order aberrations (RMS-HOA) and total RMS values.
In pseudophakic eyes, the adaptive optics Hartmann-Shack device demonstrated acceptable repeatability for measurement of sphere and 3rd and 4th order HOAs with higher variability for astigmatism measurements, especially in eyes with a prior history of corneal refractive surgery. .
评估在有和无先前角膜屈光手术史的人工晶状体眼,使用组合自适应光学视觉模拟器和哈特曼-夏克波前像差仪获得的波前像差的内 次重复性。
对每个人的每只眼进行 3 次连续测量。记录 4.5mm 瞳孔的第 5 个 Zernike 阶的总眼像差。通过计算个体内标准差(S)、重复性限(R)和组内相关系数(ICC)来评估重复性。进行向量分析以评估扫描之间的散光变异性。
该研究纳入了 32 名正常个体和 24 名有屈光手术史的个体。在正常眼和有先前屈光手术史的眼中,球镜的 S 值分别为 0.155 和 0.176 屈光度(D),柱镜的 S 值分别为 0.184 和 0.265 D。所有 3 阶项的 S 值在正常眼中为 0.037 至 0.047 µm,在有先前屈光手术史的眼中为 0.044 至 0.063 µm。正常眼中初级球差的 S 值为 0.020 µm,有先前屈光手术史的眼中为 0.026 µm。测量散光的 ICC 值变化较大(ICC = 0.751 和 0.879)。然而,两组对 RMS 高阶像差(RMS-HOA)和总 RMS 值的测量都表现出良好的可重复性(ICC > 0.9)。
在人工晶状体眼中,自适应光学哈特曼-夏克设备对球镜和 3 阶和 4 阶像差的测量具有可接受的重复性,散光测量的变异性较大,尤其是在有先前角膜屈光手术史的眼中。