Gabriel Claudia, Klaproth Oliver Klaus, Titke Christine, Baumeister Martin, Bühren Jens, Kohnen Thomas
From the Department of Ophthalmology, Goethe-University, Frankfurt, Germany.
From the Department of Ophthalmology, Goethe-University, Frankfurt, Germany.
J Cataract Refract Surg. 2015 Apr;41(4):806-11. doi: 10.1016/j.jcrs.2014.07.037.
To evaluate the repeatability of corneal first-surface aberrations and aberrometric measurements of a combined aberrometer-topographer (iTrace Visual Function Analyzer) for 5 accommodative stimuli.
Department of Ophthalmology, Goethe-University, Frankfurt, Germany.
Prospective case series.
Eyes of young adults with low to moderate myopia or hyperopia were examined in 2 measurement series comprising 3 corneal topography measurements and wavefront measurements focusing on targets at 5 distances (0.4, 0.5, 0.6, 1.0, and 4.0 m). Repeatability was analyzed using the standard deviation (SD) within 6 measurements and the coefficient of repeatability (CoR) by Bland-Altman analysis of the means of the 2 measurement series.
Fifty eyes (50 subjects) were evaluated. The lowest calculated mean SD/CoR measurements using the aberrometer function at 5 target distances were 0.07 μm/0.14 μm for total ocular aberrations root mean square (RMS), 0.07 μm/0.16 μm for lower-order aberrations (LOAs) RMS, 0.03 μm/0.66 μm for higher-order aberrations (HOAs) RMS, 0.10 diopter (D)/0.22 D for sphere, 0.08 D/0.13 D for cylinder, and 0.09 D/0.21 D for the spherical equivalent (SE). Repeatability of the total RMS, LOAs RMS, sphere, and SE measurements improved with increasing target distance (P < .01 for aberrometric values; P < .03 for refractive values). Measurement reproducibility using the corneal topographer showed mean SDs/CoRs of 0.19 μm/0.41 μm for total RMS, 0.19 μm/0.39 μm for LOAs RMS, and 0.08 μm/0.23 μm for HOAs RMS.
The aberrometer showed satisfactory results for measurement repeatability at all target distances and declining repeatability with increasing accommodation effort.
No author has a financial or proprietary interest in any material or method mentioned.
评估组合式像差仪-地形图仪(iTrace视觉功能分析仪)针对5种调节刺激的角膜前表面像差及像差测量的可重复性。
德国法兰克福歌德大学眼科。
前瞻性病例系列。
对低至中度近视或远视的年轻成年人的眼睛进行2个测量系列的检查,包括3次角膜地形图测量和针对5个距离(0.4、0.5、0.6、1.0和4.0米)目标的波前测量。使用6次测量内的标准差(SD)以及通过对2个测量系列的均值进行Bland-Altman分析得出的重复性系数(CoR)来分析可重复性。
评估了50只眼(50名受试者)。在5个目标距离下使用像差仪功能计算得出的最低平均SD/CoR测量值为:总眼像差均方根(RMS)为0.07μm/0.14μm,低阶像差(LOA)RMS为0.07μm/0.16μm,高阶像差(HOA)RMS为0.03μm/0.66μm,球镜为0.10屈光度(D)/0.22D,柱镜为0.08D/0.13D,等效球镜(SE)为0.09D/0.21D。总RMS、LOA RMS、球镜和SE测量的可重复性随目标距离增加而提高(像差值P < .01;屈光值P < .03)。使用角膜地形图仪的测量再现性显示,总RMS的平均SD/CoR为0.19μm/0.41μm,LOA RMS为0.19μm/0.39μm,HOA RMS为0.08μm/0.23μm。