Piquet J, Isabey D, Chang H K, Harf A
Am Rev Respir Dis. 1985 Jul;132(1):104-8. doi: 10.1164/arrd.1985.132.1.104.
In order to identify the useful range of frequencies and the effect of lung volume on gas exchange during high-frequency ventilation, particularly during high-frequency body surface oscillation (HFBSO), we studied 12 normal, anesthetized, and paralyzed adult rabbits in 2 groups at 1, 3, 5, 8, 12, and 16 Hz in random order. The rabbits were placed in a body box, and a "bias" flow system different from all previously reported systems was used. For a given animal at a given frequency, the oscillation magnitude was adjusted to ensure normocapnia, defined as PaCO2 equal to 40 +/- 2 mmHg, for at least 5 min. For the first group (n = 6) with a mean tracheal pressure of 0.5 cmH2O, the necessary tidal volumes (mean +/- SD) were 3.7 +/- 0.2, 1.9 +/- 0.2, and 1.5 +/- 0.1 ml/kg at 1, 3, and 5 Hz, respectively. Further increases in frequency resulted in only small decreases in the required tidal volumes: 1.4 +/- 0.1, 1.2 +/- 0.1, and 1.2 +/- 0.1 ml/kg at 8, 12, and 16 Hz, respectively. Arterial PO2 values were very similar at all 6 frequencies, with a mean of 78.6 +/- 3.3 mmHg in this group. For the second group (n = 6) with a mean tracheal pressure of 5 cmH2O, arterial PO2 values were again the same at all frequencies applied but were significantly higher (95.5 +/- 2.6 mmHg) than in the first group. No significant difference was observed in the tidal volumes required to maintain normocapnia between the 2 groups, i.e., at the 2 mean tracheal pressures studied.(ABSTRACT TRUNCATED AT 250 WORDS)
为了确定高频通气期间,尤其是高频体表振荡(HFBSO)期间的有用频率范围以及肺容积对气体交换的影响,我们将12只正常、麻醉且瘫痪的成年兔随机分为2组,按照1、3、5、8、12和16Hz的顺序进行研究。将兔子置于一个体箱中,并使用了一种与之前所有报道系统都不同的“偏置”气流系统。对于给定频率下的给定动物,调整振荡幅度以确保至少5分钟内维持正常碳酸血症,即动脉血二氧化碳分压(PaCO2)等于40±2mmHg。对于第一组(n = 6),平均气管压力为0.5cmH2O,在1、3和5Hz时,维持正常碳酸血症所需的潮气量(平均值±标准差)分别为3.7±0.2、1.9±0.2和1.5±0.1ml/kg。频率进一步增加时,所需潮气量仅略有下降:在8、12和16Hz时分别为1.4±0.1、1.2±0.1和1.2±0.1ml/kg。该组所有6个频率下的动脉血氧分压(PO2)值非常相似,平均值为78.6±3.3mmHg。对于第二组(n = 6),平均气管压力为5cmH2O,所有应用频率下的动脉PO2值同样相同,但明显高于第一组(95.5±2.6mmHg)。在两组之间,即在所研究的两个平均气管压力下,维持正常碳酸血症所需的潮气量未观察到显著差异。(摘要截短于250字)