Suppr超能文献

一种新型心脏导管室屏蔽装置

A Novel Shielding Device for Cardiac Cath Labs.

机构信息

University of Alabama at Birmingham, 1720 University Blvd., Birmingham, AL 35294.

University of Alabama at Birmingham, 806 Wells Ave SE, Huntsville, AL 35801.

出版信息

Health Phys. 2025 Jan 1;128(1):52-59. doi: 10.1097/HP.0000000000001890. Epub 2024 Sep 19.

Abstract

This research evaluates the effectiveness of a large specialized cardiac catheterization laboratory shielding device (SCCLSD) placed perpendicular to the patient compared to traditional shielding methods in reducing occupational exposure to scattered x rays, contributing to the ongoing enhancement of radiation safety in the cardiac catheterization laboratory (CCL) setting. An experimental setup involving an anthropomorphic phantom on the catheterization table simulated radiation scatter from a patient. Measurements were taken systematically at various grid points and heights in the CCL using a Fluke 451P ion chamber while mimicking a real interventional scenario. In-air peak exposure rates were analyzed at head, chest, and waist heights in the anteroposterior (AP) position. Results demonstrated that the SCCLSD provided a superior radiation shadow and effective whole-body radiation exposure reduction compared to conventional shielding devices. Considering that conventional shielding requires staff to wear lead aprons, an effective dose equivalent correction factor was applied for exposure measurements without the SCCLSD. Even after the correction factor, the SCCLSD continued outperforming lead aprons and offered whole-body protection, including the head and arms, which is typically neglected with conventional shielding. The SCCLSD also reduces exposure to the eyes, aligning with lower occupational exposure recommendations from ICRP and NCRP. However, proper CCL staff positioning is important in maximizing the effectiveness of the SCCLSD. Future research avenues may explore exposure rates at different C-arm angles to more completely assess the SCCLSD's impact on occupational exposure.

摘要

本研究评估了一种大型专用心脏导管室屏蔽装置(SCCLSD)与传统屏蔽方法相比,在减少职业性散射线暴露方面的有效性,该装置垂直于患者放置,有助于持续提高心脏导管室(CCL)的辐射安全性。实验设置包括在导管台上的人体模型,模拟来自患者的辐射散射。使用 Fluke 451P 离子室在 CCL 中以各种网格点和高度系统地进行测量,同时模拟真实的介入场景。在前-后(AP)位置,在头部、胸部和腰部高度分析空气中的峰值暴露率。结果表明,SCCLSD 提供了优越的辐射阴影和有效的全身辐射暴露减少,与传统屏蔽装置相比。考虑到传统屏蔽需要工作人员佩戴铅围裙,对于没有 SCCLSD 的暴露测量应用了有效剂量当量校正因子。即使在校正因子之后,SCCLSD 仍然优于铅围裙,并提供全身保护,包括头部和手臂,这通常被传统屏蔽所忽略。SCCLSD 还降低了眼睛的暴露,符合 ICRP 和 NCRP 的较低职业暴露建议。然而,正确的 CCL 工作人员定位对于最大限度地提高 SCCLSD 的效果很重要。未来的研究可能会探索不同 C 臂角度的暴露率,以更全面地评估 SCCLSD 对职业暴露的影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验