Russell L, Johnson T M, Lawrence Y, Reichelt B, Vanderloo N, Cufari M, Buschmann B I, Dannhoff S, DeVault A, Doeg E, Evans T, Foo B C, Frankel R, Kunimune J H, Pearcy J A, Vargas J, Gatu Johnson M, Frenje J
Massachusetts Institute of Technology, Plasma Science and Fusion Center, Cambridge, Massachusetts 02139, USA.
Rev Sci Instrum. 2024 Oct 1;95(10). doi: 10.1063/5.0219479.
This paper reports on investigations on the impact of higher neutron fluences on the detection efficiency of protons with CR-39, a charged particle track detector. CR-39 is widely used as a diagnostic for inertial fusion applications and is an integral component of numerous particle diagnostics at the OMEGA laser facility and National Ignition Facility. As experiments continue to produce higher and higher yields, existing diagnostics are impacted by higher particle fluences than they were originally designed for. This paper presents data from experiments measuring proton signal on pieces of CR-39 with different levels of neutron fluence with two different etch times. The experiments show a decrease in signal recovery with increased neutron fluence, which is exacerbated at longer etch times. At 3 h etch time, data suggest a 17% ± 7% signal loss at 1.3 × 105 neutron-induced tracks per cm2 and a 67% ± 21% loss at 6 h etch time. Careful signal isolation techniques can recover most of the proton tracks even with moderate neutron fluence.
本文报道了关于更高中子注量对使用CR - 39(一种带电粒子径迹探测器)探测质子的效率影响的研究。CR - 39被广泛用作惯性聚变应用的诊断工具,并且是欧米茄激光装置和国家点火装置众多粒子诊断的一个组成部分。随着实验持续产生越来越高的产额,现有的诊断工具受到比其最初设计时更高的粒子注量的影响。本文展示了在两种不同蚀刻时间下,对具有不同中子注量水平的CR - 39片上质子信号进行测量的实验数据。实验表明,随着中子注量增加,信号恢复率降低,在较长蚀刻时间下这种情况会加剧。在蚀刻时间为3小时时,数据表明在每平方厘米1.3×10⁵个中子诱发径迹的情况下,信号损失为17%±7%,在蚀刻时间为6小时时信号损失为67%±21%。即使在中子注量适中的情况下,仔细的信号隔离技术也能恢复大部分质子径迹。