Suppr超能文献

鞘管尺寸是否重要:各种连续流动内窥镜的流量和压力的台架比较。

Does sheath size matter: benchtop comparison of flow and pressure across variety of continuous flow endoscopes.

机构信息

Department of Urology, University of Minnesota, 420 Delaware St. Se. MMC 394, Minneapolis, MN, 55455, USA.

出版信息

World J Urol. 2024 Oct 10;42(1):573. doi: 10.1007/s00345-024-05270-0.

Abstract

INTRODUCTION

Laser enucleation utilizes purpose-built endoscopes for laser stabilization and continuous flow. No evaluation has been done with respect to flow or intravesical pressure with these scopes. We sought to evaluate the effect different endoscopes and sheath sizes on irrigation outflow and intravesical pressure.

METHODS

Using a benchtop model using a silicone bladder model, five outer/inner sheath combinations were assessed: Storz 28/26Fr, Storz 26/26Fr, Wolf 26/24Fr, Wolf 26/22Fr, and Wolf 24/22Fr. A urodynamics pressure transducer was inserted alongside the scope for bladder pressure measurement and outflow from scope to drain was measured using uroflowmetry device. Four 1-minute trials were recorded for each sheath and the steady state flow and pressure was recorded.

RESULTS

The Storz 28 F outer sheath and 26 F inner sheath had the highest outflow (12.4 ± 0.5 mL/s, p < 0.01). The Wolf 24 F outer and 22 F inner had the lowest outflow (7.0 ± 0.0 mL/s, p < 0.01). The steady state bladder pressure was the lowest in the Storz 28/26 (1.5 ± 1.7 cm H2O, p < 0.01)) and the greatest in the Storz 26/26 (24.2 ± 1.9 cm H2O, p < 0.01).

CONCLUSION

The Storz 28/26 combination had best outflow rate and lowest intravesical pressures in our benchtop study. Flow rates generally decreased with smaller sheath sizes and steady state bladder pressures increased as the difference between the outflow and inflow sheath size narrowed. These findings provide initial parameters that could guide sheath selection in future to optimize visualization and success of voiding trials.

摘要

简介

激光前列腺剜除术利用特制的内窥镜进行激光稳定和持续流动。尚未对这些内窥镜的流量或膀胱内压力进行评估。我们旨在评估不同内窥镜和鞘管尺寸对冲洗流出和膀胱内压力的影响。

方法

使用硅胶膀胱模型的台式模型,评估了五种外/内鞘组合:STORZ 28/26Fr、STORZ 26/26Fr、Wolf 26/24Fr、Wolf 26/22Fr 和 Wolf 24/22Fr。将尿动力学压力传感器插入内窥镜旁边以测量膀胱压力,并用尿流率计测量从内窥镜到排水管的流出量。每个鞘管记录了 4 次 1 分钟的试验,记录稳定状态的流量和压力。

结果

STORZ 28F 外鞘和 26F 内鞘的流出量最高(12.4±0.5mL/s,p<0.01)。Wolf 24F 外鞘和 22F 内鞘的流出量最低(7.0±0.0mL/s,p<0.01)。STORZ 28/26 组的膀胱压力最低(1.5±1.7cm H2O,p<0.01),而 STORZ 26/26 组的膀胱压力最高(24.2±1.9cm H2O,p<0.01)。

结论

在我们的台式研究中,STORZ 28/26 组合具有最佳的流出率和最低的膀胱内压力。随着鞘管尺寸的减小,流量普遍下降,而流出鞘管和流入鞘管之间的尺寸差异缩小,稳定状态膀胱压力增加。这些发现为未来的鞘管选择提供了初步参数,以优化可视化和排空试验的成功率。