数字与传统胸部影像:胶片数字射线摄影系统的观察者表现

Digital and conventional chest images: observer performance with Film Digital Radiography System.

作者信息

Goodman L R, Foley W D, Wilson C R, Rimm A A, Lawson T L

出版信息

Radiology. 1986 Jan;158(1):27-33. doi: 10.1148/radiology.158.1.3940392.

Abstract

The Film Digital Radiography System (FilmDRS) is a device with a laser optical film digitizer, 2,000 X 2,000 X 12-bit memory, and a 1,000-line video display. To evaluate the adequacy of this device for general radiography of the chest, four readers independently analyzed both radiographs and the corresponding video display of the digitized chest images of 150 patients, consisting of 100 images of abnormalities and 50 normal images. The overall results indicate equal sensitivity for the two systems. The FilmDRS, with interactive windowing, proved superior in the detection of hilar and mediastinal disease. X-ray film was superior in allowing detection of hyperlucent states. There was equivalent sensitivity for other disease categories. Superior specificity was achieved with conventional radiographs.

摘要

胶片数字射线摄影系统(FilmDRS)是一种配备激光光学胶片数字化仪、2000×2000×12位内存以及1000行视频显示器的设备。为评估该设备用于胸部常规射线摄影的适用性,四位阅片者独立分析了150例患者的射线照片及数字化胸部图像的相应视频显示,其中包括100张异常图像和50张正常图像。总体结果表明两种系统的敏感性相同。具备交互式窗口功能的FilmDRS在检测肺门和纵隔疾病方面表现更优。X线胶片在检测透亮状态方面更具优势。对于其他疾病类别,两者的敏感性相当。传统射线照片具有更高的特异性。