Chotas H G, Ravin C E
Department of Radiology, Thoracic Imaging Research Division, Duke University Medical Center, Box 3302 (Rm 139, Bryan Research Bldg), Durham, NC 27710, USA.
Radiology. 2001 Mar;218(3):679-82. doi: 10.1148/radiology.218.3.r01mr27679.
To evaluate and compare human observer performance in a contrast-detail test by using postprocessed hard-copy images from a digital chest radiography system and conventional screen-film radiographs.
The digital radiography system is based on a large-area flat-panel x-ray detector with a structured cesium iodide scintillator layer and an amorphous silicon thin-film transistor array for image readout. Images of a contrast-detail phantom were acquired at two exposure levels by using two standard thoracic screen-film systems and the digital system at matched dose. By using images of the phantom processed with standard chest image postprocessing techniques, a four-alternative forced-choice observer perception study was performed, and the number of detectable test signals (disk-shaped objects 0.3-4.0 mm in diameter) was determined for each image type.
On average, observers detected more test signals on digital images than on screen-film radiographs at all diameters up to 2.0 mm and an equivalent number at larger diameters. Test signals with lower inherent subject contrast were detected more readily on digital images than on screen-film images, even when x-ray exposure levels for the digital system were reduced by 20%.
Observer performance in a contrast-detail detection task can be improved by using images acquired with the flat-panel digital chest radiography system as compared with those acquired with state-of-the-art screen-film combinations.
通过使用数字胸部射线摄影系统的后处理硬拷贝图像和传统的屏-片射线照片,评估并比较人类观察者在对比细节测试中的表现。
数字射线摄影系统基于一个大面积平板X射线探测器,该探测器具有结构化碘化铯闪烁体层和用于图像读出的非晶硅薄膜晶体管阵列。通过使用两个标准胸部屏-片系统和数字系统在匹配剂量下,以两种曝光水平采集对比细节体模的图像。使用标准胸部图像后处理技术处理体模图像,进行四选一强制选择观察者感知研究,并确定每种图像类型可检测到的测试信号(直径为0.3 - 4.0毫米的盘状物体)数量。
平均而言,观察者在所有直径达2.0毫米的数字图像上检测到的测试信号比在屏-片射线照片上更多,在较大直径时检测到的数量相当。即使数字系统的X射线曝光水平降低20%,在数字图像上也比在屏-片图像上更容易检测到固有对比度较低的测试信号。
与使用先进的屏-片组合采集的图像相比,使用平板数字胸部射线摄影系统采集的图像可提高观察者在对比细节检测任务中的表现。