Suppr超能文献

用于乳腺癌成像的具有可穿戴乳房光学接口的MRI引导近红外光谱断层扫描(MRg-NIRST)成像系统

MRI-Guided Near-Infrared Spectroscopic Tomography (MRg-NIRST) Imaging System With Wearable Breast Optical Interface for Breast Cancer Imaging.

作者信息

Zhao Mengyang, Cao Xu, Feng Jinchao, Zhou Mingwei, Wei Chengpu, diFlorio-Alexander Roberta, Pogue Brian W, Jiang Shudong, Paulsen Keith D

出版信息

IEEE Trans Biomed Eng. 2025 Mar;72(3):899-908. doi: 10.1109/TBME.2024.3479081. Epub 2025 Feb 20.

Abstract

OBJECTIVE

To develop a novel Magnetic Resonance Imaging (MRI)-guided Near-Infrared Spectroscopic Tomography (MRg-NIRST) imaging system with an MRI-compatible breast optical interface for breast imaging.

METHODS

The breast interface consists of eight flexible opto-electronic circuit strips, each equipped with six photodetectors and six side-firing fiber-probes. Concurrent MRI and NIRST data were acquired from a total of 2,304 source-detector positions at six wavelengths, enabling 3D MRg-NIRST image reconstruction of the entire breast. The system was validated through a series of phantom and normal subject studies.

RESULTS

Reconstructed images of phantoms with inclusions ranging 10-25 mm in diameter showed errors in the estimated inclusion diameter and contrast of total hemoglobin (HbT) within the inclusion relative to the background were ranged in [-7%, 12%] and [7%, 28%], respectively. HbT estimates from reconstructed images of nine normal subjects ranged between 8.0-l25.2 μM, align with previous imaging studies.

CONCLUSION

Results from both phantom and normal subject studies indicate that this system has the potential to be easily integrated into clinical practice for acquiring 3D MRg-NIRST images of the entire breast.

SIGNIFICANCE

The flexibility of the wearable breast optical interface, along with the increased number of sources and detectors, has improved the optical accessibility for breasts of various sizes, shapes, and tumor locations. 3D MRg-NIRST image reconstruction, based on optical data collected from multiple source-detector layers across the entire breast, demonstrates that MRg-NIRST is ready to be tested clinically for its potential to enhance breast cancer detection alongside MRI.

摘要

目的

开发一种新型的磁共振成像(MRI)引导的近红外光谱断层扫描(MRg-NIRST)成像系统,该系统具有与MRI兼容的乳房光学接口,用于乳房成像。

方法

乳房接口由八个柔性光电电路条组成,每个电路条配备六个光电探测器和六个侧面发射光纤探头。在六个波长下,从总共2304个源探测器位置同时采集MRI和NIRST数据,从而实现整个乳房的三维MRg-NIRST图像重建。该系统通过一系列体模和正常受试者研究进行了验证。

结果

对直径在10-25毫米范围内包含物的体模进行重建图像显示,估计的包含物直径存在误差,且包含物内总血红蛋白(HbT)相对于背景的对比度误差分别在[-7%,12%]和[7%,28%]范围内。九名正常受试者重建图像的HbT估计值在8.0-125.2μM之间,与先前的成像研究结果一致。

结论

体模和正常受试者研究结果均表明,该系统有潜力轻松整合到临床实践中,用于获取整个乳房的三维MRg-NIRST图像。

意义

可穿戴乳房光学接口的灵活性,以及增加的源和探测器数量,提高了对各种大小、形状和肿瘤位置乳房的光学可达性。基于从整个乳房多个源探测器层收集的光学数据进行的三维MRg-NIRST图像重建表明,MRg-NIRST已准备好在临床上进行测试,以评估其与MRI一起增强乳腺癌检测的潜力。

相似文献

本文引用的文献

5
Breast cancer early detection: A phased approach to implementation.乳腺癌早期检测:分阶段实施方法。
Cancer. 2020 May 15;126 Suppl 10(Suppl 10):2379-2393. doi: 10.1002/cncr.32887.
8
Breast Cancer Treatment: A Review.乳腺癌治疗:综述。
JAMA. 2019 Jan 22;321(3):288-300. doi: 10.1001/jama.2018.19323.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验