Faculty of Chemistry, Department of Environmental Chemistry, University of Lodz, 163/165 Pomorska Str., 90-236, Łódź, Poland.
Doctoral School of Exact and Natural Sciences, University of Lodz, 12/16 Banacha Str., 90-237, Łódź, Poland.
Sci Rep. 2024 Oct 18;14(1):24425. doi: 10.1038/s41598-024-75760-z.
The report presents first high performance liquid chromatography-tandem mass spectrometry (HPLC-MS/MS) based method for the determination of plasma 2-(3-hydroxy-5-phosphonooxymethyl-2-methyl-4-pyridyl)-1,3-thiazolidine-4-carboxylic acid (HPPTCA), an adduct of cysteine and active form of vitamin B6 pyridoxal 5'-phosphate. The assay employs 4-deoxypyridoxine (4-DPD) as an internal standard. Sample preparation procedure primarily involves acetonitrile (ACN) extraction of HPPTCA from plasma proteins, sample deproteinization by ultrafiltration, and dilution of ultrafiltrate with mobile phase prior to chromatographic analysis. The chromatographic separations of HPPTCA and 4-DPD are achieved within 6 min at 20 °C on X-Bridge Glycan BEH Amide (100 × 2.1 mm, 2.5 μm) column using gradient elution. The eluent consists of 0.1% formic acid in a mixture of solvent A (water and ACN (95:5, v: v)) and solvent B (water and ACN (5:95, v: v)) delivered at a flow rate of 0.3 mL/min. The assay linearity was observed within 0.25-10 µmol/L in plasma. The limit of quantification was found to be 0.25 µmol/L. The method was successfully applied to plasma samples delivered by apparently healthy donors showing that the HPLC-MS/MS assay is suitable for human plasma screening. The presence of HPPTCA was confirmed in eleven of fifteen study samples. The HPPTCA concentration ranged from 0.55 to 8.39 µmol/L.
本报告介绍了一种基于高效液相色谱-串联质谱(HPLC-MS/MS)的方法,用于测定血浆中的 2-(3-羟基-5-膦酸氧基甲基-2-甲基-4-吡啶基)-1,3-噻唑烷-4-羧酸(HPPCA),这是半胱氨酸与维生素 B6 活性形式吡哆醛 5'-磷酸的加合物。该测定方法采用 4-脱氧吡哆醇(4-DPD)作为内标。样品制备过程主要包括用乙腈(ACN)从血浆蛋白中提取 HPPCA,用超滤法对样品进行去蛋白处理,然后用流动相稀释超滤物,再进行色谱分析。在 20°C 下,使用 X-Bridge Glycan BEH Amide(100×2.1mm,2.5μm)柱,在 6 分钟内可实现 HPPCA 和 4-DPD 的色谱分离,采用梯度洗脱。洗脱液由 0.1%甲酸在溶剂 A(水和 ACN(95:5,v:v))和溶剂 B(水和 ACN(5:95,v:v))的混合物中组成,流速为 0.3mL/min。在血浆中,该测定方法的线性范围为 0.25-10µmol/L。定量下限为 0.25µmol/L。该方法成功应用于来自健康供体的血浆样品,表明 HPLC-MS/MS 测定法适用于人血浆筛查。在十五个研究样本中有十一个样本中确认存在 HPPCA。HPPCA 浓度范围为 0.55-8.39µmol/L。