Suppr超能文献

流式细胞术试剂滴定的威力。

The Power of Reagent Titration in Flow Cytometry.

机构信息

Scientific Commercialization, Cytek Biosciences, Inc., 47215 Lakeview Boulevard, Fremont, CA 94538, USA.

Customer Education, Cytek Biosciences, Inc., 47215 Lakeview Boulevard, Fremont, CA 94538, USA.

出版信息

Cells. 2024 Oct 11;13(20):1677. doi: 10.3390/cells13201677.

Abstract

Flow cytometry facilitates the detection of multiple cell parameters simultaneously with a high level of resolution and throughput, enabling in-depth immunological evaluations. High data resolution in flow cytometry depends on multiple factors, including the concentration of reagents used in the staining protocol, and reagent validation and titration should be the first step in any assay optimization. Titration is the process of finding the concentration of the reagent that best resolves a positive signal from the background, with the saturation of all binding sites, and minimal antibody excess. The titration process involves the evaluation of serial reagent dilutions in cells expressing the antigen target for the tested antibody. The concentration of antibody that provides the highest signal to noise ratio is calculated by plotting the percentage of positive cells and the intensity of the fluorescence of the stained cells with respect to the negative events, in a concentration-response curve. The determination of the optimal antibody concentration is necessary to ensure reliable and reproducible results and is required for each sample type, reagent clone and lot, as well as the methods used for cell collection, staining, and storage conditions. If the antibody dilution is too low, the signal will be too weak to be accurately determined, leading to suboptimal data resolution, high variability across measurements, and the underestimation of the frequency of cells expressing a specific marker. The use of excess antibodies could lead to non-specific binding, reagent misuse, and detector overloading with the signal off scale and higher spillover spreading. In this publication, we summarized the titration fundamentals and best practices, and evaluated the impact of using a different instrument, sample, staining, acquisition, and analysis conditions in the selection of the optimal titer and population resolution.

摘要

流式细胞术可以同时以高分辨率和高通量检测多个细胞参数,从而实现深入的免疫学评估。流式细胞术的数据分辨率取决于多个因素,包括染色方案中使用的试剂浓度,试剂的验证和滴定应该是任何测定优化的第一步。滴定是找到最佳试剂浓度的过程,该浓度可以从背景中最佳地分辨出阳性信号,同时使所有结合位点饱和,并使抗体过量最小化。滴定过程涉及在表达测试抗体的抗原靶标的细胞中评估一系列试剂稀释度。通过绘制阳性细胞的百分比和染色细胞的荧光强度相对于阴性事件的浓度-反应曲线,计算提供最高信噪比的抗体浓度。确定最佳抗体浓度对于确保可靠和可重复的结果是必要的,并且对于每个样本类型、试剂克隆和批次以及用于细胞收集、染色和储存条件的方法都是必需的。如果抗体稀释度过低,则信号将太弱而无法准确确定,从而导致数据分辨率不佳、测量之间的变异性高以及低估表达特定标记的细胞的频率。使用过量的抗体可能导致非特异性结合、试剂滥用以及信号超出范围和更高的溢出扩散导致探测器过载。在本出版物中,我们总结了滴定的基本原理和最佳实践,并评估了使用不同仪器、样本、染色、采集和分析条件对最佳滴定度和群体分辨率选择的影响。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/3850/11506663/5c3e9ab6c63f/cells-13-01677-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验