Suppr超能文献

使用环形阵列脉冲场消融导管优化能量传递的方法和技术。

Methods and techniques to optimize energy delivery using the circular array pulsed field ablation catheter.

作者信息

Mountantonakis Stavros, Beccarino Nicholas, Abrams Mark, Sharma Nikhil, Skipitaris Nicholas, Bernstein Neil, Jia Kelly, Bhasin Kabir, Kanda Takashi, Kordalis Athanasios, Tsiachris Dimitris, Coleman Kristie

机构信息

Northwell Cardiovascular Institute, Center for Arrhythmias, New Hyde Park, New York; Department of Cardiology, Division of Electrophysiology, Lenox Hill Hospital, New York, New York.

Northwell Cardiovascular Institute, Center for Arrhythmias, New Hyde Park, New York; Department of Cardiology, Division of Electrophysiology, Lenox Hill Hospital, New York, New York.

出版信息

Heart Rhythm. 2025 Jul;22(7):e74-e84. doi: 10.1016/j.hrthm.2024.10.045. Epub 2024 Oct 24.

Abstract

Pulsed field ablation (PFA) is a novel method of cardiac ablation in which electrical fields are used to create microscopic pores in the cardiomyocyte cell membrane, resulting in cell death. Unlike traditional thermal radiofrequency and cryoablation technologies, PFA is cardiomyocyte preferential, reducing the risk of collateral damage to the esophagus and phrenic nerve. However, achieving durable lesions with PFA is dependent on the proximity to the tissue and presently approved systems do not provide contact force sensing. The PulseSelect PFA system is the first Food and Drug Administration-approved/CE-marked/Pharmaceuticals and Medical Devices Agency-approved system for PFA of atrial fibrillation; however, there is no consensus on workflow and best practices. We present here real-world experience from our centers in the United States, Europe, and Japan, and propose techniques and methods for incorporating fluoroscopy, electroanatomic mapping, and intracardiac echocardiography to assure optimal lesion delivery and predictability.

摘要

脉冲场消融(PFA)是一种新型的心脏消融方法,该方法利用电场在心肌细胞膜上形成微小孔隙,从而导致细胞死亡。与传统的热射频和冷冻消融技术不同,PFA对心肌细胞具有选择性,降低了对食管和膈神经造成附带损伤的风险。然而,使用PFA实现持久的损伤取决于与组织的接近程度,并且目前已获批准的系统不具备接触力传感功能。PulseSelect PFA系统是首个获得美国食品药品监督管理局批准/欧洲合格认证/日本药品和医疗器械局批准用于房颤PFA的系统;然而,在工作流程和最佳实践方面尚未达成共识。我们在此展示了来自美国、欧洲和日本中心的实际经验,并提出了结合荧光透视、电解剖标测和心腔内超声心动图的技术和方法,以确保实现最佳的损伤递送和可预测性。