BACKGROUND/AIM: In the pencil beam scanning carbon-ion radiotherapy, spot positions are arranged in three dimensions throughout the entire target region. Therefore, dose deviations can occur due to spot position errors in the target. However, performing three-dimensional measurements for routine patient-specific quality assurance (PSQA) is difficult because a simple measurement method has not been established. This study aimed to establish a three-dimensional dose verification method using beam delivery log data.
Pencil beam dose distributions in water were generated through Monte Carlo (MC) calculations. Treatment beam dose distribution was calculated by superposing the pencil beam dose distributions, considering given spot positions and monitor units (referred to as semi-MC, SMC). The aim of this study was to perform gamma analysis (GA) using dose distributions of log data-based SMC instead of measured dose for PSQA. To verify SMC, the SMC depth-dose curves were compared with the measured dose. To assess the equivalence between the proposed and measurement-based methods, pass rates of two-dimensional GA were compared. Furthermore, a three-dimensional GA was performed to investigate clinically suitable criteria.
The SMC dose curves were consistent with measured dose, with a deviation <5%. In two-dimensional GA, pass rates for the proposed method were generally lower than those for measurement-based method. The results of the three-dimensional GA indicated that the proposed method, with criteria of 3%-3 mm and 3%-2 mm, had capabilities comparable to the measurement-based method.
The developed three-dimensional log data-based PSQA method with criteria of 3%-3 mm and 3%-2 mm is clinically applicable.
背景/目的:在铅笔束扫描碳离子放疗中,点位置在整个靶区的三维空间中排列。因此,由于靶区中的点位置误差,可能会出现剂量偏差。然而,由于尚未建立简单的测量方法,因此对常规的患者特定质量保证(PSQA)进行三维测量具有一定难度。本研究旨在建立一种使用束流日志数据进行三维剂量验证的方法。
通过蒙特卡罗(MC)计算生成水的铅笔束剂量分布。通过叠加铅笔束剂量分布,考虑给定的点位置和监测单位(称为半 MC,SMC),计算治疗束剂量分布。本研究的目的是使用基于日志数据的 SMC 的剂量分布而不是实测剂量进行伽马分析(GA),以进行 PSQA。为了验证 SMC,将 SMC 深度剂量曲线与实测剂量进行比较。为了评估所提出的方法和基于测量的方法之间的等效性,比较了二维 GA 的通过率。此外,还进行了三维 GA 以研究临床适用的标准。
SMC 剂量曲线与实测剂量一致,偏差<5%。在二维 GA 中,所提出方法的通过率普遍低于基于测量的方法。三维 GA 的结果表明,所提出的方法具有 3%-3mm 和 3%-2mm 的标准,与基于测量的方法具有相当的能力。
所开发的具有 3%-3mm 和 3%-2mm 标准的基于日志数据的三维 PSQA 方法在临床上是可行的。