Suppr超能文献

调节 25 号和 27 号内界膜镊夹持力和刚度的因素。

Factors regulating the gripping force and stiffness of 25- and 27-gauge internal limiting membrane forceps.

机构信息

Division of Ophthalmology, Department of Surgery, Kobe University Graduate School of Medicine, Kobe, Japan.

Department of Ophthalmology, Kansai Medical University, Hirakata, Japan.

出版信息

PLoS One. 2024 Nov 5;19(11):e0310419. doi: 10.1371/journal.pone.0310419. eCollection 2024.

Abstract

This study aimed to identify the factors affecting the gripping force and stiffness of 25-gauge and 27-gauge (25G and 27G, respectively) internal limiting membrane (ILM) forceps and to compare the effect of these factors on various ILM forceps manufactured by different companies. This study evaluated 25G and 27G ILM forceps with two different types of tip shapes, Eckardt and Maxgrip, manufactured by Alcon (A), DORC (B), VitreQ (C), and Katalyst (D). The gripping force was defined as the force required to move the ILM forceps away from a thin paper by pulling the paper. Shaft stiffness was determined by measuring the shaft displacement under a known force. Multiple regression analysis revealed that the gripping force showed significant correlations with the gauge (P<0.001), type of shaft tip (Eckardt/Maxgrip) (P<0.001), and contact area of the tip (P<0.001). The shaft stiffness showed significant correlations with the gauge (P<0.001), length of the base (P<0.001), thickness of the metal of the shaft (P = 0.05), and lumen area of the shaft (P = 0.01). The gripping force and shaft stiffness differed for each product. Thus, vitreoretinal surgeons must select the appropriate type of ILM forceps based on their characteristics.

摘要

本研究旨在确定影响 25G 和 27G(分别为 25G 和 27G)内界膜(ILM)镊夹持力和刚度的因素,并比较这些因素对不同公司制造的各种 ILM 镊的影响。本研究评估了具有两种不同尖端形状(Eckardt 和 Maxgrip)的 25G 和 27G ILM 镊,这些镊由 Alcon(A)、DORC(B)、VitreQ(C)和 Katalyst(D)制造。夹持力定义为通过拉动纸张使 ILM 镊从薄纸上移动所需的力。轴刚度通过测量在已知力下轴的位移来确定。多元回归分析显示,夹持力与规格(P<0.001)、轴尖端类型(Eckardt/Maxgrip)(P<0.001)和尖端接触面积(P<0.001)显著相关。轴刚度与规格(P<0.001)、基部长度(P<0.001)、轴金属厚度(P=0.05)和轴内腔面积(P=0.01)显著相关。每种产品的夹持力和轴刚度都不同。因此,玻璃体视网膜外科医生必须根据其特点选择合适类型的 ILM 镊。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/761c/11537382/bb052613aa6f/pone.0310419.g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验