Yang JaeWon, Dasari Suhas P, Chansky Howard A, Hernandez Nicholas M
Department of Orthopaedic Surgery, University of Washington, Seattle, WA.
Arthroplast Today. 2024 Oct 18;30:101543. doi: 10.1016/j.artd.2024.101543. eCollection 2024 Dec.
Dual-mobility bearings are being increasingly utilized in total hip arthroplasty. Contemporary modular designs often feature inner cobalt-chromium liners that are seated in outer titanium acetabular shells. However, mating of these 2 dissimilar metals may lead to complications. We present a case report of a patient who was found to have osteolysis and acetabular protrusio due to an adverse local tissue reaction from corrosion between a cobalt-chromium liner and titanium acetabular shell that appeared to be well-seated clinically and radiographically. This case suggests that implant failure and associated adverse local tissue reactions may occur even in seemingly well-seated modular dual-mobility implants.
双动轴承在全髋关节置换术中的应用越来越广泛。当代的模块化设计通常具有内部钴铬衬垫,其安装在外部钛髋臼壳中。然而,这两种不同金属的配对可能会导致并发症。我们报告一例患者,该患者因钴铬衬垫与钛髋臼壳之间的腐蚀引起的局部不良组织反应,出现了骨溶解和髋臼内陷,而该假体在临床和影像学上看起来位置良好。该病例表明,即使在看似位置良好的模块化双动植入物中,也可能发生植入物失败及相关的局部不良组织反应。