手术能量设备热效应评估:离体研究。

Evaluation of thermal effects of surgical energy devices: ex vivo study.

机构信息

Department of Nephro-urology, Nagoya City University Graduate School of Medical Sciences, 1, Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya, 467-8601, Japan.

Department of Urology, Nagoya City University Midori Municipal Hospital, Nagoya, Japan.

出版信息

Sci Rep. 2024 Nov 9;14(1):27365. doi: 10.1038/s41598-024-78624-8.

Abstract

This study evaluated the direct and indirect thermal effects of various surgical energy devices using an ex-vivo model. Two types of three devices were evaluated: ENSEAL™ X1 Curved Jaw Tissue Sealer (X1) and ENSEAL™ G2 Curved Tissue Sealer (G2) as vessel sealing systems (VSSs), and HARMONIC HD1000i Shears (HA) as an ultrasonic activating device (USAD). Each device was activated once under DRY or WET conditions. The tissue's maximum temperature (MT), steam MT surrounding the activation site, and steam spread area (SSA) were measured. Under WET conditions, the median MT of a porcine common carotid artery at 1 mm from the activation site by X1, G2, and HA were 84.4, 83.3, and 50.5 °C, respectively. The direct thermal effect of HA was the lowest among the three devices. VSSs showed higher tissue MT under WET conditions compared with DRY conditions. Conversely, USAD showed the opposite trend. G2 demonstrated a significantly higher MT than X1 and HA (P < 0.05). A significant decrease in SSA was observed with decreasing grasping range. In conclusion, VSSs generated higher temperatures than USAD, especially under WET conditions. Surgeons should consider minimizing thermal effects by creating DRY conditions or performing gradual incisions when using VSS devices.

摘要

本研究采用离体模型评估了各种手术能量设备的直接和间接热效应。评估了两种类型的三种设备:ENSEAL™ X1 弯形组织封口器(X1)和 ENSEAL™ G2 弯形组织封口器(G2)作为血管密封系统(VSS),以及 HARMONIC HD1000i 剪(HA)作为超声激活装置(USAD)。每种设备在 DRY 或 WET 条件下激活一次。测量了组织的最高温度(MT)、激活部位周围的蒸汽 MT 和蒸汽扩散面积(SSA)。在 WET 条件下,X1、G2 和 HA 在距激活部位 1mm 的猪颈总动脉处的 MT 中位数分别为 84.4、83.3 和 50.5°C。三种设备中,HA 的直接热效应最低。VSS 在 WET 条件下的组织 MT 高于 DRY 条件。相反,USAD 则呈现相反的趋势。G2 的 MT 明显高于 X1 和 HA(P<0.05)。SSA 随夹持范围的减小而显著减小。总之,VSS 产生的温度高于 USAD,尤其是在 WET 条件下。当使用 VSS 设备时,外科医生应考虑通过创造 DRY 条件或进行逐步切口来最大程度地减少热效应。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/a136/11550423/2887ecb61593/41598_2024_78624_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验