An internal fixator for posterior application to short segments of the thoracic, lumbar, or lumbosacral spine. Design and testing.

作者信息

Krag M H, Beynnon B D, Pope M H, Frymoyer J W, Haugh L D, Weaver D L

出版信息

Clin Orthop Relat Res. 1986 Feb(203):75-98.

PMID:3956000
Abstract

A new spinal implant has been designed and biomechanical testing completed, intended for application to "short-segment" spinal defects such as disc degeneration, fracture, spondylolisthesis, or tumor. Major improvements over currently available devices include: only 2-3 vertebrae are spanned, not 5-7 as with Harrington rods; true three-dimensional fixation is achieved, preventing such problems as hook or rod dislocation; three-dimensional adjustment is easily accomplished, allowing fracture or spondylolisthesis reduction to be readily performed; attachment to vertebrae is by means of transpedicular screws eliminating deliberate encroachment into the spinal canal, such as Luque wires or Harrington hooks; no special alignment between screws is needed (such as with holes or slots in a plate), allowing screw placement to fully conform to anatomic structures; and laminectomy sites and lumbosacral junction are readily instrumented. Background investigations presented here for design of this device include: CT-defined pedicle morphometry showing that screws may be larger than those currently used; effect of pitch, minor diameter, and tooth profile on screw pull-out strength; mechanical testing of a compact, three-dimensionally adjustable, strong, nonloosening articulating clamp; and establishing of the relationship between depth of penetration and strength of fixation of transpeduncular screws.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验