Suppr超能文献

卡姆兰计技术说明;一种手动操作的测量唾液拉丝度的装置。

A technical note on the Kamranmeter; a manually operated device for measuring salivary spinnbarkeit.

机构信息

Department of Oral Biochemistry, Academic Centre for Dentistry Amsterdam (ACTA), University of Amsterdam and VU University Amsterdam, Amsterdam, The Netherlands.

Department of Periodontology, Academic Centre for Dentistry Amsterdam (ACTA), University of Amsterdam and VU University Amsterdam, Amsterdam, The Netherlands.

出版信息

Clin Oral Investig. 2024 Nov 22;28(12):651. doi: 10.1007/s00784-024-06047-w.

Abstract

OBJECTIVE

The objective was to develop a manually operatable, non-electric device to measure salivary spinnbarkeit for research, clinical and educational purposes.

MATERIALS AND METHODS

A newly developed device, named the Kamranmeter, was built and evaluated in a pilot study, comparing the spinnbarkeit of unstimulated and chew-stimulated saliva samples from healthy volunteers. The Neva Meter, which operates on electrical resistance, was used as a reference standard.

RESULTS

The study found no statistically significant differences between the two devices when measuring the spinnbarkeit of both unstimulated and stimulated saliva samples from healthy volunteers.

CONCLUSIONS

We developed a hand-operated device to analyse salivary spinnbarkeit. The Kamranmeter performs comparably to the Neva Meter in measuring spinnbarkeit in saliva obtained from healthy volunteers.

CLINICAL RELEVANCE

The Kamranmeter has potential applications in diagnostic tests involving saliva, such as assessing spinnbarkeit to evaluate conditions like caries and dry mouth. Moreover, it holds promise for broader medical uses, including fertility assessments, where spinnbarkeit is a key indicator of cervical mucus quality. Its ease of handling is expected to make it well-suited for research, clinical, and educational applications.

摘要

目的

研发一种手动操作、非电动的设备,用于研究、临床和教育目的,以测量唾液拉丝度。

材料和方法

在一项初步研究中构建并评估了一种新开发的设备,命名为 Kamranmeter,比较了健康志愿者的未刺激和咀嚼刺激唾液样本的拉丝度。使用基于电阻的 Neva Meter 作为参考标准。

结果

研究发现,当测量健康志愿者的未刺激和刺激唾液样本的拉丝度时,两种设备之间没有统计学上的显著差异。

结论

我们开发了一种手动操作的设备来分析唾液拉丝度。Kamranmeter 在测量来自健康志愿者的唾液拉丝度方面与 Neva Meter 表现相当。

临床相关性

Kamranmeter 在涉及唾液的诊断测试中具有应用潜力,例如评估拉丝度以评估龋齿和口干等情况。此外,它有望在更广泛的医学用途中得到应用,包括生育能力评估,其中拉丝度是宫颈黏液质量的关键指标。其易于操作预计使其非常适合研究、临床和教育应用。