To evaluate and compare the clinical outcomes of rotationally asymmetric multifocal intraocular lenses (IOLs) in a mix-and-match approach or bilateral implantation.
In this prospective study, patients were split into two groups based on the types of IOLs implanted. The mix group received mix-and-match implantation of a LENTIS Mplus LS-313 +1.50D IOL in the dominant eye and a LENTIS Mplus LS-313 +3.0D IOL in the fellow eye. The bilateral group received bilateral implantation of LENTIS Mplus LS-313 +3.00D IOLs. A three-month follow-up was performed after the surgery. Binocular visual acuity at various distances, subjective refraction and defocus curves were assessed. Subjective questionnaires included the overall satisfaction, a self-assessment of visual quality, spectacle independence, and presence of photic phenomena were completed.
The study was completed by 52 patients in total (25 mix-and-match, 27 bilateral). The mean binocular uncorrected intermediate visual acuity in the mix group and the bilateral group were 0.10 ± 0.12 logMAR and 0.23 ± 0.14 logMAR, respectively, demonstrating a significant difference (P = 0.005). The mix group performed better in intermediate vision (range: -1.00 to -2.00 D) than the bilateral group (all P< 0.05), according to binocular defocus curves. About 92.0% of patients in the mix group did not need spectacles for intermediate distance vision compared to 59.26% in the bilateral group (P = 0.0001).
The mix-and-match implantation of the +1.50D and +3.00D rotationally asymmetric multifocal IOLs provided good binocular visual outcomes and resulted in better intermediate vision than bilateral implantation of +3.00D IOLs.
评估和比较旋转不对称多焦点人工晶状体(IOL)采用混合匹配法或双侧植入的临床效果。
在这项前瞻性研究中,根据植入的IOL类型将患者分为两组。混合组在优势眼植入LENTIS Mplus LS - 313 +1.50D IOL,在对侧眼植入LENTIS Mplus LS - 313 +3.0D IOL。双侧组双侧植入LENTIS Mplus LS - 313 +3.00D IOL。术后进行三个月的随访。评估不同距离的双眼视力、主观验光和散焦曲线。完成主观问卷,包括总体满意度、视觉质量自我评估、无需眼镜情况和光现象情况。
共有52例患者完成研究(25例混合匹配,27例双侧植入)。混合组和双侧组的平均双眼未矫正中间视力分别为0.10±0.12 logMAR和0.23±0.14 logMAR,差异有统计学意义(P = 0.005)。根据双眼散焦曲线,混合组在中间视力(范围:-1.00至-2.00 D)方面比双侧组表现更好(所有P<0.05)。混合组约92.0%的患者中间距离视力无需眼镜,而双侧组为59.26%(P = 0.0001)。
+1.50D和+3.00D旋转不对称多焦点IOL的混合匹配植入提供了良好的双眼视觉效果,并且比双侧植入+3.00D IOL产生更好的中间视力。