Numerous intraoral scanners (IOS) devices are currently used for intraoral impressions in prosthodontics. It is crucial to assess the accuracy and precision of these devices for clinical use.
This in vitro study aimed to assess and compare the accuracy and precision of six IOS in the all-on-4 treatment concept. Trueness evaluation involved aligning the data from six dental scanners with the reference model. Precision analysis encompassed aligning the data from all six scanners within their respective groups.
An edentulous maxillary model was utilized, with four implants placed at 12, 15, 22, and 25 teeth positions, simulating the all-on-4 approach using resin acrylic. Following the placement of scanbodies, each of the six IOSs (Primescan, Trios 3, Trios 4, Trios 5, Virtuo Vivo, and Medit i 700) performed eight scans of the model. An industrial scanner was employed for the control group. Data alignment and comparison were executed using the CloudCompare software (v2.11.3, General Public License of Telecom ParisTech, Paris, France). Statistical scrutiny encompassed the Shapiro-Wilk, Levene's, and Games-Howell tests.
Among the scanners, Primescan exhibited the highest trueness (35.75 ± 26.08 μm), whereas the Medit i700 demonstrated superior precision (0.163 μm).
IOS can be used to make dental impressions within the all-on-4 concept. More comprehensive and clinical studies are needed on this subject.
目前,众多口内扫描仪(IOS)设备用于口腔修复学中的口内印模制取。评估这些设备在临床应用中的准确性和精确性至关重要。
本体外研究旨在评估和比较六款IOS在全口四颗种植体治疗理念中的准确性和精确性。准确性评估包括将来自六款牙科扫描仪的数据与参考模型对齐。精确性分析包括在各自组内对齐来自所有六款扫描仪的数据。
使用无牙上颌模型,在12、15、22和25牙位植入四颗种植体,采用树脂丙烯酸模拟全口四颗种植体方法。放置扫描体后,六款IOS(Primescan、Trios 3、Trios 4、Trios 5、Virtuo Vivo和Medit i 700)中的每一款对模型进行八次扫描。对照组使用工业扫描仪。使用CloudCompare软件(v2.11.3,法国巴黎电信巴黎高等矿业学校通用公共许可证)进行数据对齐和比较。统计分析包括Shapiro-Wilk检验、Levene检验和Games-Howell检验。
在这些扫描仪中,Primescan表现出最高的准确性(35.75±26.08μm),而Medit i700表现出更高的精确性(0.163μm)。
IOS可用于全口四颗种植体理念下制取牙印模。关于该主题需要进行更全面的临床研究。