Tytler J A, Seeley H F
Anaesthesia. 1986 Mar;41(3):302-5. doi: 10.1111/j.1365-2044.1986.tb12793.x.
The accuracy of the Nellcor N-101 pulse oximeter has been evaluated in adult patients receiving general anaesthesia or intensive care. Readings obtained noninvasively with this instrument were compared with measurements made on arterial blood using a Radiometer OSM2 oximeter. The pulse oximeter was easy to use and within the range tested (70-100 percent saturation of haemoglobin with oxygen) the readings were within I digit of the values obtained by in vitro measurement.