Spiss C K, Mauritz W, Zadrobilek E, Draxler V
Anaesthesist. 1985 Aug;34(8):405-8.
Non-invasive measurements of oxygen saturation are attractive because they provide continuous information which may result in improved patient care. We evaluated a new finger pulse oximeter as a measure of arterial oxygen saturation in critically ill patients with respiratory distress. In 12 patients studied, 133 simultaneous comparisons between in vivo (Pulse-Oximeter, Nellcor, Hayward, California) and in vitro (IL CO-Oximeter 282) oxygen-haemoglobin saturation were made. Linear regression analysis of these pooled data yielded on excellent correlation (r = 0.97, p less than 0.001, 0.3% accuracy). The results demonstrate that non-invasive oxygen monitoring of patients with respiratory failure is feasible and can reliably detect potential life-threating arterial oxygen desaturation.
无创测量血氧饱和度很有吸引力,因为它能提供持续信息,这可能会改善患者护理。我们评估了一种新型手指脉搏血氧仪,用于测量呼吸窘迫的重症患者的动脉血氧饱和度。在研究的12名患者中,对体内(Nellcor脉搏血氧仪,加利福尼亚州海沃德市)和体外(IL CO-血氧仪282)的氧合血红蛋白饱和度进行了133次同步比较。对这些汇总数据进行线性回归分析,得出了极佳的相关性(r = 0.97,p < 0.001,准确度为0.3%)。结果表明,对呼吸衰竭患者进行无创氧监测是可行且能可靠检测出潜在威胁生命的动脉血氧饱和度降低情况。