[脉搏血氧饱和度测量作为一种连续的、非侵入性监测方法。两种仪器的比较]

[Pulse oximetry as a continuous, noninvasive monitoring procedure. Comparison of 2 instruments].

作者信息

Altemeyer K H, Mayer J, Berg-Seiter S, Fösel T

出版信息

Anaesthesist. 1986 Jan;35(1):43-5.

PMID:3963352
Abstract

Pulse oximetry allows continuous registration of the arterial oxygen saturation by using the light absorption in a wave range between 600 and 1,000 nm. In addition the peripheral pulse is determined by a plethysmographic method. Two new devices, the Biox III pulse oximeter and the Nellcor pulse oximeter were compared for the following three items: 1. The accuracy in comparison with a simultaneously determined arterial oxygen saturation. Concerning the Biox III instrument, the correlation coefficient resulted in r = 0.90, for the Nellcor device in r = 0.92. 2. The onset time for registering a suddenly appearing hypoxia: For the Biox III oximeter we found a time lag between 6 and 42 s (median 21 s), for the Nellcor Oximeter between 12-39 s (median 21 s). 3. The pulse frequency corresponded very well with the simultaneously registered heart rate of the ECG. Our results indicate that pulse oximetry--a continuous and noninvasive method--is a reliable monitor for oxygen saturation, and provides trend informations about circulation, both of which are particularly appropriate for patients breathing spontaneously.

摘要

脉搏血氧饱和度仪通过利用600至1000纳米波长范围内的光吸收来持续记录动脉血氧饱和度。此外,外周脉搏通过容积描记法测定。对两种新设备,即Biox III脉搏血氧饱和度仪和Nellcor脉搏血氧饱和度仪进行了以下三项比较:1. 与同时测定的动脉血氧饱和度相比的准确性。对于Biox III仪器,相关系数r = 0.90,对于Nellcor设备,r = 0.92。2. 记录突然出现的低氧血症的起始时间:对于Biox III血氧饱和度仪,我们发现滞后时间在6至42秒之间(中位数为21秒),对于Nellcor血氧饱和度仪,在12至39秒之间(中位数为21秒)。3. 脉搏频率与同时记录的心电图心率非常吻合。我们的结果表明,脉搏血氧饱和度测定法——一种连续且无创的方法——是一种可靠的血氧饱和度监测手段,并能提供有关循环的趋势信息,这两者对于自主呼吸的患者尤为适用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验