Suppr超能文献

在ECT期间,MECTA脑电图癫痫发作持续时间测量的评分者间信度较差。

Poor interrater reliability of MECTA EEG seizure duration measurement during ECT.

作者信息

Ries R K

出版信息

Biol Psychiatry. 1985 Jan;20(1):94-8. doi: 10.1016/0006-3223(85)90140-4.

Abstract

Seizure duration assessed from the MECTA EEG (mEEG) tape was not found to be a reliable measurement across five independent examiners. Although all examiners agreed that an electrical seizure was demonstrated on the mEEG, the examiners were markedly different in their duration assessments (p less than 0.0001). Criteria are needed for mEEG measurement and interpretation.

摘要

通过MECTA脑电图(mEEG)磁带评估的癫痫发作持续时间,在五名独立检查人员中被发现并非可靠的测量指标。尽管所有检查人员都一致认为mEEG上显示有癫痫电发作,但他们在发作持续时间的评估上存在显著差异(p小于0.0001)。mEEG的测量和解读需要标准。