Achieving reduced radiation exposure with maintained fluoroscopy effectiveness using ultralow-dose settings in cryoballoon ablation.

作者信息

Kaneshiro Takashi, Murota Sadahiro, Nehashi Takeshi, Nodera Minoru, Yamada Shinya, Ikeda Masamitsu, Takeishi Yasuchika

机构信息

Department of Cardiovascular Medicine Fukushima Medical University Fukushima Japan.

Department of Arrhythmia and Cardiac Pacing Fukushima Medical University Fukushima Japan.

出版信息

J Arrhythm. 2024 Nov 6;40(6):1400-1407. doi: 10.1002/joa3.13179. eCollection 2024 Dec.

Abstract

BACKGROUND AND AIMS

Optimization of fluoroscopic image quality for reducing radiation exposure in cryoballoon pulmonary vein isolation (CB-PVI) has not yet been fully investigated. Therefore, we tried to compare the radiation doses among three different X-ray system settings.

METHODS

Consecutive 148 patients scheduled for their first CB-PVI were prospectively enrolled: low dose with the use of an anti-scatter grid for the first 51 patients (LD + G group), low dose without an anti-scatter grid for the subsequent 46 patients (LD-G group), and ultralow dose (ULD group) with an anti-scatter grid for the remaining 51 patients. We compared the radiation doses required to complete CB-PVI procedures among the groups. There were 27 patients for whom CB-PVI was performed without cine acquisition, but with fluoroscopy only, and the radiation doses were also compared.

RESULTS

The median procedure time and fluoroscopy time were 119 and 35.5 min, respectively, with no significant differences among the groups. The median cumulative air Kerma (AK) decreased in both the LD-G group (71.8 mGy,  < .001) and the ULD group (73.0 mGy,  < .001), compared to the LD + G group (145.0 mGy). Among 27 patients who underwent CB-PVI without cine acquisition, the median cumulative AK further decreased in both the LD-G group (31.4 mGy,  < .05) and the ULD group (22.7 mGy,  < .01), compared to the LD + G group (64.6 mGy).

CONCLUSION

Using an ULD X-ray setting and avoiding cine acquisition, we can reduce radiation exposure, while ensuring the necessary fluoroscopy time for the CB-PVI procedure.

摘要

背景与目的

冷冻球囊肺静脉隔离术(CB-PVI)中用于减少辐射暴露的透视图像质量优化尚未得到充分研究。因此,我们试图比较三种不同X射线系统设置下的辐射剂量。

方法

前瞻性纳入连续148例计划首次行CB-PVI的患者:前51例患者使用防散射格栅的低剂量组(LD+G组),随后46例患者不使用防散射格栅的低剂量组(LD-G组),其余51例患者使用防散射格栅的超低剂量组(ULD组)。我们比较了各组完成CB-PVI手术所需的辐射剂量。有27例患者仅进行透视而未采集电影图像进行CB-PVI,也比较了他们的辐射剂量。

结果

各组的中位手术时间和透视时间分别为119分钟和35.5分钟,差异无统计学意义。与LD+G组(145.0 mGy)相比,LD-G组(71.8 mGy,P<0.001)和ULD组(7