Suppr超能文献

在机械通气期间使用声学氦分析仪和微处理器快速测量绝对肺容积。

Use of an acoustic helium analyzer and microprocessor for rapid measurement of absolute lung volume during mechanical ventilation.

作者信息

Gillespie D J

出版信息

Crit Care Med. 1985 Feb;13(2):118-21. doi: 10.1097/00003246-198502000-00014.

Abstract

A technique developed to measure functional residual capacity during mechanical ventilation uses an acoustic helium analyzer and microprocessor to perform open-circuit helium washout. The acoustic helium analyzer has acceptable linearity up to a helium concentration of 50% and is stable. The microprocessor combines output from the acoustic analyzer and a ventilator with an expiratory flow signal, to compute lung volume rapidly during mechanical ventilation. In a sample of ten patients with various lung problems, open-circuit helium washout gave results similar to those of a closed-circuit helium-dilution technique during assist control or intermittent mandatory ventilation modes, with or without positive end-expiratory pressure. This open-circuit system has an advantage over other techniques because it is small and portable and can measure lung volume quickly without complicated maneuvers.

摘要

一种在机械通气期间测量功能残气量的技术,使用声学氦分析仪和微处理器来进行开路氦气冲洗。该声学氦分析仪在氦气浓度高达50%时具有可接受的线性度且性能稳定。微处理器将声学分析仪和呼吸机的输出与呼气流量信号相结合,以便在机械通气期间快速计算肺容积。在10名患有各种肺部问题的患者样本中,在辅助控制或间歇强制通气模式下,无论有无呼气末正压,开路氦气冲洗得到的结果与闭路氦气稀释技术的结果相似。这种开路系统比其他技术具有优势,因为它体积小、便于携带,并且无需复杂操作就能快速测量肺容积。