拇外翻锯骨模型中经皮关节外横向跖骨颈截骨术不同固定方法的生物力学评估

Biomechanical Evaluation of Different Fixation Methods for Percutaneous Extracapsular Transverse Cervical Metatarsal Osteotomy in a Hallux Valgus Sawbone Model.

作者信息

Wagner Pablo, Wagner Emilio, Pinochet Miguel, Salinas David, Lam Peter

机构信息

Orthopaedic department, Clinica Alemana de Santiago-Universidad del Desarrollo, Santiago, Chile.

Orthopaedic department, Hospital Militar de Santiago-Universidad de los Andes, Santiago, Chile.

出版信息

Foot Ankle Orthop. 2024 Dec 13;9(4):24730114241303768. doi: 10.1177/24730114241303768. eCollection 2024 Oct.

Abstract

BACKGROUND

There is limited literature available that provide information about fixation methods for minimally invasive hallux valgus osteotomies. Our objective was to evaluate the strength of different fixation methods for a percutaneous extracapsular transverse cervical metatarsal (PTCM) osteotomy in a sawbone model.

METHODS

Thirty solid foam sawbone foot models were used. Percutaneous extracapsular transverse cervical metatarsal osteotomies were performed and fixed in a standardized fashion in 6 different groups: (1) one 4.0-mm screw; (2) 2 (one 4.0-mm and one 3.0-mm) parallel screws; (3) 2 (one 4.0-mm and one 3.0-mm) divergent screws; (4) same as group 3, but with lateral metatarsal head cortex purchase with the 4.0-mm screw; (5) same as group 4, but with two 4.0-mm screws; (6) same as group 5, but with two 3.5-mm screws. Cyclic and load to failure testing were performed applying a plantar load to the metatarsal head. The measured variables were stiffness and force needed to create deformity using a Zwick Roell Universal Testing Machine.

RESULTS

Group 1 failed as a result of rotational instability in cyclic testing. In load to failure testing, group 3 did not achieve difference compared with group 2 ( = .09) (181 vs 131 N). Group 4 was stronger ( = .02) (250 N) than groups 1-3. Group 5 did not show differences relative to group 4 ( = .1) (223 N). Group 6 was stronger only than groups 1 and 2 ( = .01) (193 N).

CONCLUSION

In this sawbones-based study, we found that the use of two 4.0-mm screws or use of a 3.0-mm screw and a 4.0-mm screw with lateral first metatarsal head cortical purchase was superior to other screw configurations.

摘要

背景

关于微创拇外翻截骨术固定方法的文献资料有限。我们的目的是在锯骨模型中评估经皮关节外横向跖骨颈(PTCM)截骨术不同固定方法的强度。

方法

使用30个实心泡沫锯骨足模型。进行经皮关节外横向跖骨颈截骨术,并以标准化方式固定于6个不同组:(1)一枚4.0毫米螺钉;(2)2枚(一枚4.0毫米和一枚3.0毫米)平行螺钉;(3)2枚(一枚4.0毫米和一枚3.0毫米)发散螺钉;(4)与第3组相同,但4.0毫米螺钉固定于外侧跖骨头皮质;(5)与第4组相同,但使用两枚4.0毫米螺钉;(6)与第5组相同,但使用两枚3.5毫米螺钉。对跖骨头施加足底负荷进行循环和破坏载荷测试。使用Zwick Roell万能试验机测量的变量为刚度和造成畸形所需的力。

结果

第1组在循环测试中因旋转不稳定而失败。在破坏载荷测试中,第3组与第2组相比无差异(P = 0.09)(181对131 N)。第4组比第1 - 3组更强(P = 0.02)(250 N)。第5组与第4组相比无差异(P = 0.1)(223 N)。第6组仅比第1组和第2组更强(P = 0.01)(193 N)。

结论

在这项基于锯骨的研究中,我们发现使用两枚4.0毫米螺钉或使用一枚3.0毫米螺钉和一枚4.0毫米螺钉并固定于外侧第一跖骨头皮质优于其他螺钉配置。