Materials (Basel). 2024 Nov 21;17(23):5689. doi: 10.3390/ma17235689.
A differential scanning calorimeter (DSC) is widely used for measuring the thermal properties of phase-change materials (PCMs). Optimizing test conditions based on material characteristics is essential for accurate results. This study investigates the effects of experimental parameters, including sample mass, heating rate, measurement modes, and atmosphere flow rate, on the phase-change enthalpy and phase-change temperature results. The findings indicate that variations in sample mass and heating rate lead to significant changes in phase-change temperatures, while an increase in purge gas flow rate reduces the phase-change enthalpy of the PCM. Based on the measurements, this study optimizes the DSC parameters and provides a reference for the accurate measurement of paraffin-based phase-change materials.
差示扫描量热仪(DSC)被广泛用于测量相变材料(PCM)的热性能。根据材料特性优化测试条件对于获得准确结果至关重要。本研究考察了实验参数,包括样品质量、加热速率、测量模式和气氛流速,对相变焓和相变温度结果的影响。研究结果表明,样品质量和加热速率的变化会导致相变温度发生显著变化,而吹扫气体流速的增加会降低PCM的相变焓。基于这些测量,本研究优化了DSC参数,并为基于石蜡的相变材料的准确测量提供了参考。