Suppr超能文献

在健康对照组中,从光刺激到瞳孔最大收缩的时间存在显著的个体内差异。

There Is Significant Within-Subject Variation in the Time from Light Stimulus to Maximum Pupil Constriction Among Healthy Controls.

作者信息

Kamal Abdulkadir, Kim Yohan, Salter Amber, Gunna Shripal, Nairon Emerson B, Olson DaiWai M

机构信息

Department of Neurology, The University of Texas Southwestern Medical Center (UTSW), Dallas, TX 75390, USA.

出版信息

J Clin Med. 2024 Dec 6;13(23):7451. doi: 10.3390/jcm13237451.

Abstract

: Handheld quantitative pupilometers (QPs) measure each phase of the pupillary light reflex (PLR) and provide a summary score based on these values. One phase of the PLR is the period of time from the onset of light exposure to the maximum constriction of the pupil, also known as time to maximum constriction (tMC). Although tMC has been found to vary significantly among patients with neurological injury, there are no studies reporting tMC in healthy controls. This study addresses this gap. : Subjects in this prospective observational study were healthy controls who provided paired (left and right eye) QP readings during four separate observations over the course of 2 days. The tMC was derived by determining the smallest observed pupil size during videos filmed at 30 frames per second, and we assessed within-subject variability using the coefficient of variance and intraclass correlation coefficient (ICC). : Fifty subjects provided 380 QP readings (190 left eye and 190 right eye). Subjects primarily identified as female (80%), non-Hispanic (86%), white (62%), and <40 years old (74%). The mean tMC was 1.0 (0.14) seconds (s) for the left eye and 1.0 (0.17) s for the right eye; the coefficient of variance ranged from 11.6% to 18.8% and the ICC ranged from 0.25 to 0.40. For the between-subject comparisons across the four observation periods, the left and right eye mean differences ranged from 0.001 to 0.063 and the ICC ranged from 0.12 to 0.52. : The tMC values vary significantly in healthy controls. Changes in pupil function as a clinical biomarker of intracranial pathology are not fully understood. Identifying clinical correlations of tMC variation may provide insight for the prognostication and treatment of neurocritically ill patients.

摘要

手持式定量瞳孔计(QPs)可测量瞳孔光反射(PLR)的每个阶段,并根据这些值提供一个汇总分数。PLR的一个阶段是从光照开始到瞳孔最大收缩的时间段,也称为最大收缩时间(tMC)。尽管已发现tMC在神经损伤患者中差异显著,但尚无研究报告健康对照者的tMC情况。本研究填补了这一空白。:这项前瞻性观察性研究的受试者为健康对照者,他们在2天内的4次单独观察中提供了配对(左眼和右眼)的QP读数。tMC是通过确定以每秒30帧拍摄的视频中观察到的最小瞳孔大小得出的,我们使用方差系数和组内相关系数(ICC)评估受试者内变异性。:50名受试者提供了380次QP读数(190次左眼和190次右眼)。受试者主要为女性(80%)、非西班牙裔(86%)、白人(62%)且年龄小于40岁(74%)。左眼的平均tMC为1.0(0.14)秒(s),右眼为1.0(0.17)s;方差系数范围为11.6%至18.8%,ICC范围为0.25至0.40。在四个观察期的受试者间比较中,左眼和右眼的平均差异范围为0.001至0.063,ICC范围为0.12至0.52。:健康对照者的tMC值差异显著。瞳孔功能变化作为颅内病变的临床生物标志物尚未完全明确。确定tMC变化的临床相关性可能为神经危重症患者的预后和治疗提供见解。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/e3bb/11642143/c179167eda46/jcm-13-07451-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验